Abstract
Cerenkov luminescence imaging (CLI) has been an evolutional and alternative approach of nuclear imaging in basic research. This study aimed to measure the 131I thyroid uptake of mouse using CLI for assessment of thyroid function. Quantification of 131I thyroid uptake of mice in euthyroid, hypothyroid and hyperthyroid status was performed by CLI and γ-scintigraphy at 24 hours after injection of 131I. The 131I thyroid uptake was calculated using the equation: (thyroid counts − background counts)/(counts of injected dose of 131I) × 100%. Serum T4 concentration was determined to evaluate the thyroid function. The radioactivity of 131I was linearly correlated with the CL signals in both in vitro and in vivo measurements. CLI showed a significant decrease and increase of 131I thyroid uptake in the mice in hypo- and hyperfunctioning status, respectively, and highly correlated with that measured by γ-scintigraphy. However, the percent thyroid uptake measured by CLI were one-fifth of those measured by γ-scintigraphy due to insufficient tissue penetration of CL. These results indicate that CLI, in addition to nuclear imaging, is able to image and evaluate the 131I thyroid uptake function in mice in preclinical and research settings.
Similar content being viewed by others
Introduction
Radioiodine has long been used for thyroid studies1, 2. Radioactive iodine uptake (RAIU) and thyroid scan are commonly used to evaluate patients with suspected thyroid disorders. RAIU that measures iodine metabolism in the thyroid gland is based on physiologic incorporation of radioiodide into the thyroid gland and is followed by determination of the fraction of the dose taken up in the gland over a given time period2. It has been widely used to assess the thyroid function in clinical practice. RAIU along with a thyroid scan is useful in differentiating the causes of hyperthyroidism, such as Graves’ disease, Plummer’s disease and subacute thyroiditis2. However, these nuclear medicine procedures are seldom applied on small animals in the preclinical studies due to the drawbacks of low spatial and temporal resolution of nuclear imaging and relatively high cost of instruments1, 2.
Cerenkov luminescence (CL) has been used recently in biomedical applications of imaging with clinically relevant medical isotopes3, 4. CL is a phenomenon which was first described by Pavel Alekseyevich Cherenkov in 19345. Charged particles traveling faster than the speed of light in a medium transfer their kinetic energy through interactions with the surrounding dipoles, in biological tissues mostly with water5. The randomly oriented water molecules will align with the passing super-relativistic charged particle and relax by releasing the transferred energy in the form of light5, 6. Cerenkov luminescence imaging (CLI) can be done using a sensitive camera optimized for low light condition which has a better resolution than any other nuclear imaging modality3. CLI has emerged quickly in preclinical molecular imaging with the use of various medically relevant radioisotopes, including 15O, 13N, 68Ga, 89Zr, 64Cu, 225Ac, 90Y, 131I, 124I, 18F and 74As, showing that the emitted radioactivity correlated well with the light output7,8,9,10.
With the development of optical imaging techniques and the improvement of its equipped CCD systems, the sensitivity and accuracy for CL detection has made this method highly applicable and competent in preclinical basic research setting11. In addition, advantages such as lower cost than that of nuclear imaging modalities, without setup of radiochemistry facilities, nearly identical imaging procedures and short image acquisition time of radionuclide-based optical CLI have provided an alternative and potential approach for researchers who are not majored in the nuclear imaging but need to conduct the radionuclide-based research.
The preclinical use of CLI in the study of radioiodine uptake are becoming more common in recent years. Jeong et al. successfully showed the CLI and nuclear imaging of 131I and 124I uptake by mouse thyroid gland and by NIS-expressing thyroid cancer cells12. Using 99mTc-pertechnetate, Boschi et al. demonstrated CLI of the thyroid glands and salivary glands of mice13. Spinelli et al. reported the first 131I CLI of a human thyroid gland with a therapeutic dose of 131I for the treatment of hyperthyroidism14. Hu et al. demonstrated the application of Cerenkov luminescence tomography for evaluating the 131I uptake function of mouse thyroid15. These results have shown the translational potential of CLI.
To the best of our knowledge, there is no report showing that CLI is able to quantitatively measure the thyroid uptake of radioiodide. In this study, we have approved that CLI is feasible to measure the 131I thyroid uptake in mice in different status of thyroid function.
Materials and Methods
Mouse model of hypothyroidism and hyperthyroidism
The animal manipulation and experiment procedures were reviewed and approved by the Institutional Animal Care and Use Committee of National Yang-Ming University. Six week-old Balb/c male mice, purchased from National Laboratory Animal Center (NLAC, Taiwan), were housed in a temperature-controlled room with a 12-hour light-dark cycle and given access to food and water ad libitum. For induction of hypothyroidism, the mice were orally administered with methimazole (MMI, Sigma-Aldrich), 64 mg/kg in 200 μl distilled water, via gavage for 15 consecutive days. For induction of hyperthyroidism, the mice were intramuscularly injected with 2 μg of recombinant human thyroid stimulating hormone (rhTSH, Thyrogen, thyrotropin alfa, Genzyme) in 100 μl distilled water twice on two consecutive days. The control mice were administered with the same volumes of distilled water. Measurement of serum total T4 was carried out by using a Mouse/Rat Thyroxine ELISA kit (GenWay Biotech. Inc., San Diego, CA, USA) to assess the thyroid function.
Correlation of CL signals and 131I activity in vitro and in vivo
To assess the correlation of CLI signals and radioiodide activity in vitro, 131I was serially diluted into 500, 250, 125, 63, 32 and 16 μCi (18.5, 9.3, 4.6, 2.3, 1.2 and 0.6 MBq) in 200 μl saline in Eppendorf tubes. CLI was carried out using an IVIS 50 imaging system (Perkin Elmer, Waltham, MA, USA) with a luminescence imaging setting of binning: 8, FOV: 12, f stop: 1, exposure time: 300 s. The signal of 131I-emitted CL was analyzed using Living Imaging Software (Perkin Elmer) and shown in p/s/cm2/sr.
To evaluate the correlation of injected 131I dose and CL signals in mouse thyroid, serial doses of 131I (0, 0.6, 1.2, 2.3, 4.6, 9.3 and 18.5 MBq in 200 μl saline) were intraperitoneally injected into mice followed by CLI 24 hours later. Before imaging, the mice were anesthetized by inhalation of 2% isofluorane mixed with oxygen. Quantification of CL signals from mouse thyroid was performed following the settings as described in in vitro study.
131I thyroid imaging and measurement of thyroid uptake function by γ-scintigraphy and CLI
131I thyroid imaging and measurement of thyroid uptake function were carried out on six euthyroid mice, six mice in hypothyroid status and six in hyperthyroid status. The mice were intraperitoneally injected with 18.5 MBq of 131I, followed by γ-scintigraphy and CLI 24 hours later. γ-scintigraphy was performed by placing animals prone at a distance of 6 cm under a 4-mm pinhole collimator equipped on a gamma camera (Symbia E, Siemens). Image was acquired for 30 min. CLI was performed subsequently using an IVIS 50 imaging system with the same luminescence imaging settings mentioned above. The radioactivity of the thyroid gland was obtained from the region-of-interest (ROI) drawn along 85% isocount contour of the thyroid gland over the anterior neck. The background activity was measured over the head with a ROI of the same size as the thyroid gland. The photon flux of the CL of the thyroid gland and the background was obtained using the similar method as γ scintigraphy. The percent thyroid uptake of 131I was calculated using the equation ((thyroid counts − background counts)/counts of injected 131I dose × 100%). Correction of the decay of radiotracer was done for each measurement.
Statistical analysis
The numerical data were reported as means ± standard deviation (S.D.). Student’s t test was applied for comparison of serum T4 concentration and percent 131I thyroid uptake of each group with different thyroid function status. A significant difference was considered if the p value was less than 0.05.
Ethical approval
All applicable international, national, and/or institutional guidelines for the care and use of animals were followed. This article does not describe any studies with human participants performed by any of the authors.
Results
Correlation of CLI signals and 131I activity in vitro and in vivo
131I with activity ranging from 500 to 16 μCi (18.5 to 0.6 MBq) was aliquoted in the same volume (200 μl saline), placed in Eppendorf tubes, and then imaged by CLI using the IVIS 50 system. The luminescence intensity of 131I corresponded well with the doses of radiotracer (Fig. 1A). Within a ~5 cm2 ROI, 500 μCi (18.5 MBq) and 16 μCi (0.6 MBq) of 131I resulted in signal of approximately 3 × 105 and 2 × 104 p/s/cm2/sr, respectively. A high correlation (R2 = 0.9994) was observed between the luminescence intensities and 131I activities (Fig. 1B).
To determine whether the accumulation of different doses of 131I in the thyroid gland of healthy mice can be accurately measured by CLI, doses of 131I from 0 to 500 μCi (0 to 18.5 to MBq) were intraperitoneally injected into the mice and CLI was performed 24 hours later. The results showed that the CL emitted from the thyroid gland was visually correlated with the amount of 131I activity injected (Fig. 1C). The signals from the thyroid ROI were ranging from 3 × 104 p/s/cm2/sr (18.5 MBq 131I) to 4 × 102 p/s/cm2/sr (0.6 MBq 131I). The CL signals were well correlated with the 131I doses (R2 = 0.9708) (Fig. 1D). The results suggested that the minimum dosage of 131I CLI imaging of mouse thyroid gland is 125 μCi (4.6 MBq).
In vivo 131I CLI and γ-scintigraphy of the mice in different thyroid function status
Mouse models of hypothyroidism and hyperthyroidism were established by treatment with MMI and Thyrogen, respectively. Total serum T4 was measured to confirm the successful induction of hypothyroidism and hyperthyroidism. The results showed that the serum T4 level was significantly higher in rhTSH-treated mice (~14.8 ± 3.5 μg/dl) and significantly decreased in MMI-treated mice (~2.1 ± 0.4 μg/dl), as compared to that in the euthyroid subjects (~3.7 ± 0.9 μg/dl) (p < 0.05) (Fig. 2).
The mice in hypothyroid or hyperthyroid status were intraperitoneally administrated with 500 μCi (18.5 MBq) of 131I, followed by CLI and γ-scintigraphy 24 hours later. The CL images showed obvious reduction or increase of luminescence signals in the thyroid glands in hypothyroid or hyperthyroid status, respectively, as compared to the euthyroid mice (Fig. 3). ROI analysis showed that the CL signals were 2 × 103 ~ 2 × 104 p/s/cm2/sr for the hypothyroid mice, 1.2 ~ 1.4 × 106 p/s/cm2/sr for the hyperthyroid mice and 4 ~ 7 × 105 p/s/cm2/sr for the control euthyroid mice. A similar result was observed by γ-scintigraphy, showing reduced or increased γ signal in the thyroid glands of mice in hypothyroid or hyperthyroid status, respectively (Fig. 3). The radioactivity (counts per second, cps) obtained from the ROIs of the thyroid glands were 10 ~ 50 cps for hypothyroidism, 200 ~ 300 cps for hyperthyroidism and 70 ~ 130 cps for euthyroid mice. The percent 131I uptake of the thyroid measured by CLI were 0.6 ± 0.3%, 3.0 ± 0.6% and 7.3 ± 0.6% for hypothyroid, euthyroid and hyperthyroid status, respectively. By γ-scintigraphy, the percent 131I uptakes of thyroid were 2.7 ± 1.7%, 12.9 ± 3.3% and 31.9 ± 5.3% for hypothyroid, euthyroid and hyperthyroid status, respectively (Fig. 4A and B). The value of the percent thyroid uptake for each group measured by CLI is about one-fifth of that measured by γ-scintigraphy. The percent thyroid uptake of 131I measured from the two imaging modalities were well correlated and had a high linear regression (R2 = 0.9621) (Fig. 4C), indicating the feasibility and reliability of CLI for the measurement of 131I thyroid uptake.
Discussion
In this study, we have demonstrated the feasibility of using CLI for quantification of 131I thyroid uptake function in mice. The results reveal a high linear correlation between 131I activity and the CL signals in the thyroid gland of the mice. In hypothyroid or hyperthyroid mice, CLI provides a reliable quantitative measurement of 131I thyroid uptake, which is comparable with the uptake measured by γ-scintigraphy and compatible with the thyroid function status of the mice.
The hypothyroid and hyperthyroid mice were induced by administration of MMI and rhTSH, respectively. Administration of rhTSH into normal subjects resulted in a significant increase in serum T4 at 8 h, reached to peak at 24 h and remained the high T4 level for at least 4 days16. Increased radioiodine uptake by normal subjects has also been observed at 6 and 24 hours after rhTSH treatment17. Serum T3 and T4 concentrations in euthyroid mice are strikingly increased at 6 hour after rhTSH treatment18. In the current study, the total T4 in euthyroid mice was 3.7 ± 0.9 μg/dl (n = 6), comparable with a range of 4 ~ 8 μg/dl for balb/c mice in earlier literatures19, 20 and with the data outlined by the Mouse Phenome Database at the Jackson laboratory. At 24 hours after two injections of rhTSH (totally 4 μg), serum T4 was effectively elevated by four folds (14.8 μg/dl). The effect of rhTSh on the radioiodine uptake of euthyroid mice has also been studied with 125I, which showed decreased uptake during the first 3 to 5 hours after treatment and then increased at 13 hours18. Based on these reports, 24 hours after rhTSH treatment was assumed to be the most suitable time point for the 131I imaging in mice. Notably, the mice with rhTSH stumulation showed a two-fold increase in 131I accumulation in thyroid glands as compared to that in control mice in this study. MMI has been used to treat hyperthyroidism caused by Graves’ disease for more than half a century, and is able to effectively induce hypothyroidism in euthyroid rodents21, 22. The mice were treated with ~1.6 mg (64 mg/kg) of MMI for 15 consecutive days. The serum T4 concentration (2.1 ± 0.4 μg/dl) was reduced to almost half of that in control mice (3.7 ± 0.9 μg/dl). 131I uptake by the thyroid gland measured by both CLI and γ-scintigraphy showed a decrease by one fifth of that of the euthyroid mice. In summary, hypothyroidism or hyperthyroidism of mice induced by MMI or rhTSH, respectively, is able to be monitored by 131I CLI with the same efficacy as γ-scintigraphy.
Obvious tissue attenuation of CL signal was seen in the animal experiments in this study. The percent thyroid uptake of 131I calculated from CL was about one fifth of that measured by γ-scintigraphy. The CL intensity is relatively weak and the most intense CL occurs in the spectrum of the UV and blue wavebands. UV and blue light are easily scattered and absorbed by biological tissues23. In addition, IVIS imaging system used in the current study is able to measure light signal with wavelength from 490 to 850 nm, which falls in the spectrum of green to red/near infrared light. Most CL signal cannot be detected by the optical device we used and consequently the percent thyroid uptake of 131I measured by CLI would be lower than by γ-scintigraphy. In the current study, 24 hours after administration of 18.5 MBq of 131I, the calculated accumulation of 131I in the hypofunctioning and hyperfunctioning thyroids from γ images was 0.5 and 5.9 MBq, respectively. This amount of 131I accumulation enables CLI to obtain enough photons to calculate the thyroid uptake which shows a high linear regression of correlation to the γ-scintigraphy, despite the remarkable attenuation of the CL signal. The dose of 131I calculated by the equation of minimum efficacious radiopharmaceutical dosage for scintigraphic imaging of small children and infants as (adult dosage)(child’s weight in Kg)/70 Kg, should be 0.15 μCi (0.006 MBq)24, which is about 0.03% of the administered dose of 500 μCi (18.5 MBq) in the current study. 131I has a physical half-life of 8.1 days and decays by beta emission. The major gamma emission is at 364 keV. The long half-life and the beta decay result in relatively high radiation on thyroid gland and other organs. The estimated absorbed radiation dose to the thyroid gland is 16,000 rad/mCi in the newborn, 800 rad/mCi in the adult25 and about 77,700 rad/mCi in rat26. The absorbed radiation dose to the mouse thyroid gland would be much higher. The dosage of 131I of 500 μCi (18.5 MBq) used in the current study might damage the thyroid gland and cause hypothyroidism of the animal in the long run. According to the results of the current study, we would recommend that 125 μCi (4.6 MBq) is the minimum dosage of 131I for optimal CLI of mouse thyroid gland.
In recent years, investigators have attempted to use CLI more effectively in clinical experiments by improving the image equipment. Spinelli et al. optimized the CCD detector for the near infrared (NIR) range, which contained a smaller number of Cerenkov photons, but had better tissue penetration ability. The sensitivity was improved up to 35%, and this advantage was more significant when the Cerenkov source is being imaged at a deeper location within the animal27. CCD systems with enhanced sensitivity for detecting low-light levels, such as intensified CCD (ICCD) and electron-multiplying CCD (EMCCD), have been used in recent studies14, 28, 29. These highly sensitive intensified CCD systems have recently been coupled to optical fiber allowing the demonstration of preclinical endoscopic surgery30. The results of a previous clinical study showed a good consistence with clinical PET examinations and the signals obtained by the endoscopic CLI to detect gastrointestinal malignant neoplasms31. These newly developed technology of optical imaging instruments are expected to improve both the quality of 131I imaging of thyroid gland and the measurement of 131I thyroid uptake by CLI as well.
In conclusion, we have demonstrated the feasibility of using CLI for quantitatively assessment of thyroid uptake function of radioiodide. CLI is able to quantify the 131I accumulation in the thyroid glands of mice with different functional status. The thyroid uptake of 131I measured by CLI is comparable to that obtained by γ-scintigraphy. In addition to nuclear medicine imaging modalities, CLI is useful to assess the thyroid uptake function of radioiodine in preclinical and basic research settings.
References
Hertz, S., Roberts, A. & Evans, R. D. Radioactive Iodine as an Indicator in the Study of Thyroid Physiology. Experimental Biology and Medicine 38, 510–513 (1938).
Hoffer, P. B., Gottschalk, A. & Quinn, J. I. In Diagnostic Nuclear Medicine (eds A. Gottschalk & E.J. Potchen) Ch. 20, 267–272 (Williams & Wilkins, 1976).
Das, S., Thorek, D. L. & Grimm, J. Cerenkov imaging. Adv Cancer Res 124, 213–234 (2014).
Chin, P. T. et al. Optical imaging as an expansion of nuclear medicine: Cerenkov-based luminescence vs fluorescence-based luminescence. Eur J Nucl Med Mol Imaging 40, 1283–1291 (2013).
Cerenkov, P. A. Visible emission of clean liquids by action of γ radiation. Dokl Akad Nauk SSSR 2, 451–454 (1934).
Ross, H. H. Measurement of.beta.-emitting nuclides using Cerenkov radiation. Analytical Chemistry 41, 1260–1265 (1969).
Robertson, R., Germanos, M. S., Manfredi, M. G., Smith, P. G. & Silva, M. D. Multimodal Imaging with 18F-FDG PET and Cerenkov Luminescence Imaging After MLN4924 Treatment in a Human Lymphoma Xenograft Model. Journal of Nuclear Medicine 52, 1764–1769 (2011).
Liu, H. et al. Molecular optical imaging with radioactive probes. PloS one 5, e9470 (2010).
Ruggiero, A., Holland, J. P., Lewis, J. S. & Grimm, J. Cerenkov Luminescence Imaging of Medical Isotopes. Journal of Nuclear Medicine 51, 1123–1130 (2010).
Boschi, F. et al. In vivo (1) (8)F-FDG tumour uptake measurements in small animals using Cerenkov radiation. Eur J Nucl Med Mol Imaging 38, 120–127 (2011).
Li, C., Mitchell, G. S. & Cherry, S. R. Cerenkov luminescence tomography for small-animal imaging. Optics letters 35, 1109–1111 (2010).
Jeong, S. Y. et al. Combined Cerenkov luminescence and nuclear imaging of radioiodine in the thyroid gland and thyroid cancer cells expressing sodium iodide symporter: initial feasibility study. Endocrine journal 58, 575–583 (2011).
Boschi, F. et al. Small-animal radionuclide luminescence imaging of thyroid and salivary glands with Tc99m-pertechnetate. J Biomed Opt 18, 76005 (2013).
Spinelli, A. E. et al. First human Cerenkography. J Biomed Opt 18, 20502 (2013).
Hu, Z. et al. Three-dimensional noninvasive monitoring iodine-131 uptake in the thyroid using a modified Cerenkov luminescence tomography approach. PloS one 7, e37623 (2012).
Ramirez, L., Braverman, L. E., White, B. & Emerson, C. H. Recombinant Human Thyrotropin Is a Potent Stimulator of Thyroid Function in Normal Subjects. The Journal of Clinical Endocrinology & Metabolism 82, 2836–2839 (1997).
Torres, M. S. T., Ramirez, L., Simkin, P. H., Braverman, L. E. & Emerson, C. H. Effect of Various Doses of Recombinant Human Thyrotropin on the Thyroid Radioactive Iodine Uptake and Serum Levels of Thyroid Hormones and Thyroglobulin in Normal Subjects. The Journal of Clinical Endocrinology & Metabolism 86, 1660–1664 (2001).
Colzani, R. M., Alex, S., Fang, S. L., Braverman, L. E. & Emerson, C. H. The effect of recombinant human thyrotropin (rhTSH) on thyroid function in mice and rats. Thyroid: official journal of the American Thyroid Association 8, 797–801 (1998).
Kita-Furuyama, M. et al. Dendritic cells infected with adenovirus expressing the thyrotrophin receptor induce Graves’ hyperthyroidism in BALB/c mice. Clin Exp Immunol 131, 234–240 (2003).
Maier, J. et al. Iodine deficiency activates antioxidant genes and causes DNA damage in the thyroid gland of rats and mice. Biochim Biophys Acta 1773, 990–999 (2007).
Cano-Europa, E. et al. Methimazole-induced hypothyroidism causes alteration of the REDOX environment, oxidative stress, and hepatic damage; events not caused by hypothyroidism itself. Ann Hepatol 9, 80–88 (2010).
Cano-Europa, E., Blas-Valdivia, V., Franco-Colin, M., Gallardo-Casas, C. A. & Ortiz-Butron, R. Methimazole-induced hypothyroidism causes cellular damage in the spleen, heart, liver, lung and kidney. Acta Histochem 113, 1–5 (2011).
Ma, X., Wang, J. & Cheng, Z. Cerenkov Radiation: A Multi-functional Approach for Biological Sciences. Frontiers in Physics 2 (2014).
Neel, H. L. & Becerra, D. N. In Pediatric Nuclear Medicine (eds J.H. Miller & M.J. Gelfand) 7–9 (WB Saunders and Co, 1994).
Keyes, J. W. Jr., Thrall, J. H. & Carey, J. E. Technical considerations in in vivo thyroid studies. Semin Nucl Med 8, 43–57 (1978).
Spetz, J., Rudqvist, N. & Forssell-Aronsson, E. Biodistribution and dosimetry of free 211At, 125I- and 131I- in rats. Cancer Biother Radiopharm 28, 657–664 (2013).
Spinelli, A. E. & Boschi, F. Optimizing in vivo small animal Cerenkov luminescence imaging. J Biomed Opt 17, 040506 (2012).
Carpenter, C. M. et al. Cerenkov Luminescence Endoscopy: Improved Molecular Sensitivity with β−-Emitting Radiotracers. Journal of Nuclear Medicine 55, 1905–1909 (2014).
Thorek, D. L., Riedl, C. C. & Grimm, J. Clinical Cerenkov luminescence imaging of (18)F-FDG. Journal of nuclear medicine: official publication, Society of Nuclear Medicine 55, 95–98 (2014).
Liu, H. et al. Intraoperative Imaging of Tumors Using Cerenkov Luminescence Endoscopy: A Feasibility Experimental Study. Journal of Nuclear Medicine 53, 1579–1584 (2012).
Hu, H. et al. Feasibility study of novel endoscopic Cerenkov luminescence imaging system in detecting and quantifying gastrointestinal disease: first human results. Eur Radiol 25, 1814–1822 (2015).
Acknowledgements
Dr. Ya-Ju Hsieh was supported partially by Kaohsiung Medical University “Aim for the Top Universities Grant” (KMU-TP103) and by the Ministry of Science and Technology of Taiwan (MOST 103-2314-B-037-007; MOST 104-2314-B-037-055-MY2). Dr. Chien-Chih Ke was supported by the Ministry of Science and Technology of Taiwan (MOST 105-2314-B-010-011 and MOST 106-2623-E-010 -001 -NU). Professor Ren-Shyan Liu was supported by the National Research Program for Biopharmaceuticals at the Ministry of Science and Technology of Taiwan (MOST 105-2314-B-010-034-MY2; MOST 105-2314-B-010-011), and supported by the Department of Health (MOHW106-TDU-B-211-144-003) and the Taipei Veterans General Hospital (V105C-038). We thank for technical support and Ms. Tsuey-Ling Jan for the assistance of preparing the manuscript. We thank the Taiwan Mouse Clinic (MOST 105-2325-B-001-010, MOST 106-2319-B-001-004) which is funded by the National Research Program for Biopharmaceuticals (NRPB) at the Ministry of Science and Technology (MOST) of Taiwan for technical support in PET imaging experiment.
Author information
Authors and Affiliations
Contributions
C.C.K., W.S.H. and R.S.L. conceived and designed the experiment. Z.M.H., C.C.K., Y.J.H., C.W.H., J.J.L., Y.A.C. and C.W.C. performed the experimental work. B.H.Y., W.S.H., L.H. and C.C.K. analyzed the data. C.C.K. and Y.J.H. wrote the manuscript.
Corresponding author
Ethics declarations
Competing Interests
The authors declare that they have no competing interests.
Additional information
Publisher's note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Ke, CC., He, ZM., Hsieh, YJ. et al. Quantitative Measurement of the Thyroid Uptake Function of Mouse by Cerenkov Luminescence Imaging. Sci Rep 7, 5717 (2017). https://doi.org/10.1038/s41598-017-05516-5
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-017-05516-5
This article is cited by
-
In-situ x-ray fluorescence imaging of the endogenous iodine distribution in murine thyroids
Scientific Reports (2022)
-
Biodistribution and pharmacokinetic profiles of an altered peptide ligand derived from heat-shock proteins 60 in Lewis rats
Cell Stress and Chaperones (2020)
-
Quantification of radioactivity by planar gamma-camera images, a promoted method of absorbed dose in the thyroid after iodine-131 treatment
Scientific Reports (2018)
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.