Introduction

Pars plana vitrectomy (PPV) based on 20-gauge instruments (20 G) has become and remained a widely used vitreoretinal surgical procedure and an essential therapeutic approach in many vitreoretinal diseases since 1975 [1]. This conventional 20 G PPV surgery requires the closure of the sclerotomies with sutures at the end of surgery to prevent wound leakage. In contrast, the introduction of smaller diameter instruments in the mid-2000s do not require placement of scleral sutures because of the self-sealing nature of the wounds. This surgical method is also known as microincision vitrectomy surgery (MIVS) and can be achieved with the use of 23-gauge (23 G), 25-gauge (25 G) or 27-guage (27 G) instruments, although the 27 G method is still much less widely used compared to the other two methods. MIVS is gaining popularity for the surgical treatment of several vitreoretinal diseases since it was introduced due to many potential advantages, including reduction in inflammation, patient discomfort, shorter recovery time, and in some cases, shorter duration of surgery [2].

Recently, a meta-analysis demonstrated that 23 G MIVS may carry similar risk of postoperative endophthalmitis compared to conventional 20 G vitrectomy and 25 G MIVS may carry slightly higher rate of postoperative endophthalmitis [3]. However, it is not clear if such differences appear in other complications of vitrectomy. Despite continuous advances to improve safety of both traditional 20 G PPV and MIVS, iatrogenic retinal breaks (IRBs) and postoperative retinal detachments (RDs) remain major complications of this type of surgery, usually necessitating a subsequent treatment during surgery (either cryopexy or laser) or even repeat surgery and frequently impacting visual outcomes [4]. The overall incidence of IRBs associated with 20 G vitrectomy in the published literature ranges from 1.4 to 44% [5,6,7], which appears to be somewhat higher compared to the published frequency of IRBs in MIVS, which has a range of 1.1−16% [8,9,10]. The incidence of the other major complication—postoperative RDs after 20 G vitrectomy has been reported to be between 0.8 and 9.8% [11,12,13]. While some studies have evidenced postoperative RDs following MIVS, the incidence after this type of surgery seems to be generally lower compared to IRBs, ranging from 1 to 1.6% [8, 10, 14]. However, there is still some controversy about whether MIVS carries comparable risk of IRBs and postoperative RDs compared to conventional 20 G vitrectomy and an aggregate analysis of suitable comparative studies is lacking.

Thus far, only one meta-analysis published in 2013 reported complications in MIVS compared to 20 G vitrectomy, indicating that there are no statistically significant differences between groups regarding incidence rates of choroidal detachment, vitreous hemorrhage, and postoperative RDs [2]. It included only four trials in the comparison of postoperative RDs. Furthermore, we are not aware of any meta-analysis focused on comparing the incidence of IRBs associated with both types of surgery. The availability of more recent relevant data stimulated us to undertake a new meta-analysis in order to get more informed conclusions about the incidence of these two complications in MIVS and traditional 20 G vitrectomy.

Materials and methods

Search strategy

We selected relevant studies by searching PubMed, ISI Web of Science and Embase, using the terms “pars plana vitrectomy” or “vitrectomy” or “20-gauge” and “23-gauge” or “25-gauge” or “27-gauge” or “transconjunctival sutureless vitrectomy” or “microincision vitrectomy” or “small gauge vitrectomy”, with the language restricted to English, up to July 23, 2018. Reference lists of all retrieved articles were also searched manually.

Inclusion and exclusion criteria

We included studies in this meta-analysis if the studies met the following inclusion criteria: (1) comparative cases series studies reporting the IRBs and postoperative RDs rates in all types of MIVS (23 G, 25 G, 27 G) versus 20 G vitrectomy, (2) independent prospective or retrospective studies that provided the data to estimate odds ratios (OR) with 95% confidence interval (CI). Articles with repeat data, abstracts from conferences, letters, case reports, and review articles were also excluded.

Data extraction

Two authors (GHC and WSL) selected relevant studies and extracted the data. Uncertainty was resolved by discussion. The following items were collected: first author’s name, the year of publication, study design, country of origin, participant numbers in each group, and incidence of IRBs and postoperative RDs.

Qualitative assessment

We adopted the Newcastle Ottawa Scale (NOS) to evaluate the quality of nonrandomized studies by a modified checklist [15], which has an eight-item scale based on three criteria: selection, comparability, and outcome. Studies with NOS score above 7 points were defined as high quality; as medium quality if they were between 4 and 6 points; otherwise they were poor quality. We only included the studies with NOS score above four points. The qualities of randomized clinical trials (RCTs) were assessed by Jadad scale [16], which is a 5-point scale, assigning scores for reported randomization, masking and participant withdrawals. Studies scoring less than 3 points were excluded from this study.

Statistical analysis

The Cochrane Review Manager (RevMan, software version 5.1, Copenhagen, Denmark: The Nordic Cochrane Center, The Cochrane Collaboration, 2011) was used for all statistical analyses. The OR of IRBs and postoperative RDs comparing MIVS with conventional 20 G vitrectomy were measured, and outcomes were reported with a 95% CI. Subgroup analyses based on macular surgery and different types of MIVS were conducted. Subgroup estimates by type of IRBs (sclerotomy-related retinal breaks or surgically induced posterior vitreous detachment (PVD)-related retinal breaks) were also conducted. We used the Q statistic-test to calculate heterogeneity between studies [17]. If P < 0.05, a random-effects model was applied to the data, otherwise a fixed-effects model was used. Publication bias was assessed by Begg’s rank correlation test and Egger’s linear regression test (P < 0.05 was considered representative of significant statistical publication bias) [18, 19].

Results

Overall characteristics of selected studies and quality assessment

Initially, a total of 2173 articles were identified. We excluded 2141 studies according to the exclusion criteria listed above. Hence, 32 studies were included in this meta-analysis [5,6,7,8,9,10,11,12,13,14, 20,21,22,23,24,25,26,27,28,29,30,31,32,33,34,35,36,37,38,39,40,41]. The flow diagram of the search procedure and results is summarized in Supplementary Appendix 1. Most of the studies were retrospective, except three prospective studies [20, 22, 41], and three RCTs [30, 38, 39]. According to the NOS used for quality assessment for nonrandomized studies, 14 studies had high quality scores of 7, while 15 studies had moderate quality scores of 6. The average score of all studies included in this meta-analysis was 6.48. All the RCTs fulfilled the quality criteria (3 points or more), according to the Jadad scale. In total, there were 14,373 eyes included in the meta-analysis: 6932 eyes were included in the 20 G vitrectomy group, and 7441 eyes were included in the MIVS group (23 G or 25 G). No studies allowing comparison with 27 G MIVS were identified as a result of the search. The characteristics of the studies included in the meta-analysis are summarized in Table 1.

Table 1 Characteristics of the studies included in the meta-analysis

Estimated effects of MIVS on incidence of IRBs and postoperative RDs

Twenty-six studies involving 11,275 eyes compared the rate of IRBs in 20 G vitrectomy vs. the rate in MIVS. The overall incidence of IRBs was 12.78% (771 eyes per 6033 cases) for 20 G vitrectomy, and 6.29% (330 eyes per 5242 cases) for all types of MIVS. The primary comparison clearly demonstrated that the incidence of IRBs was significantly higher in 20 G vitrectomy than in 23 G/25 G MIVS (OR = 2.22, 95% CI, 1.93−2.57, P < 0.00001) (Fig. 1a). In a subgroup analysis based on different types of MIVS (23 G or 25 G), we also found an increased risk of IRBs in 20 G vitrectomy compared to 23 G MIVS (OR = 2.19, 95% CI, 1.87−2.56, P < 0.00001) and also compared to 25 G MIVS (OR = 2.27, 95% CI, 1.62−3.18, P < 0.00001) (Fig. 1b, c). A sub-analysis in cases involving macular surgery only also demonstrated higher incidence of IRBs in 20 G vitrectomy vs. 23 G/25 G vitrectomy (OR = 2.44, 95% CI, 1.99−2.99, P < 0.00001) (Fig. 2a). In a further subgroup analysis, 15 studies involving 4986 eyes compared the rate of IRBs related to sclerotomy, and 4 studies involving 472 eyes reported IRBs associated with PVD during vitrectomy between these two groups. The result of this subgroup analysis demonstrated an increased risk of sclerotomy-related retinal breaks in 20 G vitrectomy compared with 23 G/25 G MIVS (OR = 3.73, 95% CI, 2.55−5.44, P < 0.00001) (Fig. 3a), but not for the same comparison in surgically induced PVD-related retinal breaks (OR = 1.59, 95% CI, 0.89−2.84, P = 0.12) (Fig. 3b).

Fig. 1
figure 1

Odds ratio of appearance of iatrogenic retinal breaks in microincision vitrectomy surgery with conventional 20-gauge vitrectomy. a 23-gauge or 25-guage vitrectomy, b 23-gauge vitrectomy only, c 25-gauge vitrectomy only. G gauge, MIVS microincision vitrectomy surgery

Fig. 2
figure 2

Subanalysis estimating the odds ratio of iatrogenic retinal breaks and postoperative retinal detachments in microincision vitrectomy surgery vs. conventional 20-gauge vitrectomy for macular diseases. a Iatrogenic retinal breaks, b postoperative retinal detachments. G gauge, MIVS microincision vitrectomy surgery

Fig. 3
figure 3

Subanalysis estimating the odds ratio of iatrogenic retinal breaks in microincision vitrectomy surgery vs. conventional 20-gauge vitrectomy for: a sclerotomy-related retinal breaks or b PVD-related retinal breaks. G gauge, MIVS microincision vitrectomy surgery, PVD posterior vitreous detachment

In the second part of our meta-analysis, we compared the rate of RD after surgery related to the size of instrumentation used. Eighteen studies involving 7022 eyes compared the rate of postoperative RDs in 20 G vitrectomy vs. the rate in 23 G/25 G MIVS. The overall incidence of postoperative RDs was 2.52% (76 eyes per 3010 cases) for 20 G vitrectomy vs.1.57% (63 eyes per 4012 cases) for MIVS (Table 1). The meta-analysis demonstrated a higher incidence of postoperative RDs in 20 G vitrectomy compared to that of 23 G/25 G MIVS (OR = 1.72, 95% CI, 1.21−2.46, P = 0.003) (Fig. 4a). In a subgroup analysis based on different types of MIVS (23 G or 25 G), we also found an increased risk of postoperative RDs in 20 G vitrectomy compared to 23 G MIVS (OR = 2.45, 95% CI, 1.50−4.00, P = 0.0003), but not for 25 G MIVS (OR = 1.01, 95% CI, 0.63−1.64, P = 0.96) (Fig. 4b, c). When a subanalysis including only cases of macular surgery was conducted, it also demonstrated higher incidence of RDs in 20 G vitrectomy vs. 23 G/25 G vitrectomy (OR = 1.89, 95% CI, 1.26−2.85, P = 0.002) (Fig. 2b).

Fig. 4
figure 4

Odds ratio of postoperative retinal detachments in microincision vitrectomy surgery with conventional 20-gauge vitrectomy. a 23-gauge or 25-guage vitrectomy, b 23-gauge vitrectomy only, c 25-gauge vitrectomy only. G gauge, MIVS microincision vitrectomy surgery

In studies comparing IRBs incidence rate, slightly more than half of the studies, 15 out of 26 studies (57.7%) in 20 G and 14 out of 27 studies (53.8%) in 23 G/25 G MIVS have less than 100 cases per group. Similarly, in studies reporting RDs, 10 out of 18 (55.6%) in 20 G and 9 out of 18 in 23 G/25 G MIVS (50.0%) have less than 100 cases per group. Therefore we conducted a subanalysis that only included the studies having more than 100 cases per group. The results of this subanalysis also demonstrated higher incidence of IRBs and postoperative RDs in 20 G vitrectomy than in MIVS (OR = 2.32, 95% CI, 1.97−2.72, P < 0.00001 and OR = 1.53, 95% CI, 1.01−2.32, P = 0.04 respectively) (Supplementary Appendix 2).

In addition, there were no publication bias in IRBs between the 20 G vitrectomy and 23 G/25 G MIVS (Begg’s test, P = 0.947; Egger’s test, P = 0.585) and in postoperative RDs (Begg’s test, P = 0.449; Egger’s test, P = 0.135).

Discussion

This meta-analysis included 32 relevant studies from different locations (18 from Europe, 11 from Asia and 3 from N. America), representing a wide geographic origin. It showed that the incidence rate of IRBs in 20 G vitrectomy was 12.78% (771/6033), while it was 6.29% (330/5242) in 23 G/25 G MIVS, and the incidence rate of postoperative RDs in 20 G vitrectomy was 2.52% (76/3010), while it was 1.57% (63/4012) in 23 G/25 G MIVS. The meta-analysis demonstrated that the rates of IRBs and postoperative RDs in conventional 20 G procedure were higher compared to the rates in MIVS, and similar results were obtained in a subanalysis based on macular surgery. Similar difference in incidence was shown in another subanalysis for the comparison of sclerotomy-related retinal breaks, but not for surgically induced PVD-related retinal breaks.

Although MIVS have gained considerable popularity as a preferred method for the treatment of several vitreoretinal diseases, the use of 20 G vitrectomy has retained some popularity in some parts of the world. Thus, a recent paper by Fang et al. [42]. indicates that in 2015 ~15% of all vitrectomies were done with a 20 G system at the Eye, Ear, Nose and Throat Hospital of Fudan University, Shanghai, China. A recent survey among 78 vitreoretinal surgeons from German-speaking countries indicated that 19.5% still use 20 G vitrectomy [43]. Another recent study by Tayyab et al. [44]. used equal number of patients (42/42) for 20 G vs. 23 G vitrectomy in Al Ehsan Eye Hospital, Lahore from June 2016 to March 2017. Another recent study by Ghasemi et al. [45]. used 22/20 patients for 20 G vs. 23 G vitrectomy in Tehran, Iran.

Iatrogenic retinal breaks are known as some of the most important and serious complications in vitreoretinal surgery [4]. Excessive vitreous traction on the retina is the leading cause of such breaks [46]. Several mechanisms can cause abnormal vitreoretinal traction during vitrectomy, including insertion of surgical instruments which causes traction on the adjacent vitreous base, or incarceration of vitreous in the sclerotomy site during withdrawal of instruments [47]. Iatrogenic retinal breaks located away from the entry sites are typically related to separation of the vitreous or epiretinal membranes from the retinal surface during surgery. Furthermore, intraoperative induction of PVD has been identified as an important cause of IRBs during surgery [48].

The results of the current meta-analysis clearly revealed that the incidence of IRBs associated with MIVS was significantly (~50%) lower compared to the incidence associated with 20 G vitrectomy, especially for sclerotomy-related retinal breaks. This difference may be due to several factors. One of the key factors may be the use of microcannulas, which remain within the sclerotomies for the duration of the surgery for the instruments to pass through. The cannulas may be long enough to traverse the vitreous base, enabling instruments to pass repeatedly into the eye without engaging vitreous gel around pars plana leading to less traction and fewer IRBs [27]. The higher cutting rate of a vitreous cutter may also reduce the incidence of IRBs. The 23 G/25 G MIVS cutters typically have a higher maximum cut rate of 2500/5000 cuts per minute compared with up to 1800 cuts per minute for 20 G cutters [26, 45]. Higher cut rates result in a lower volume of aspirated tissue per cut at a fixed flow, resulting in less tissue traction, which may affect break rate [26, 49]. Furthermore, the traction range of 20 G cutters is higher compared to the traction range of 23 G/25 G. Thus, pneumatic 20 G cutters had a traction range of 2.06−37.22 dynes vs. 3.85−15.38 dynes for 23 G and 5.13−27.91 dynes for 25 G and the traction range for electric 20 G cutters was even higher (3.60−41.78 dynes) [50]. This difference in traction caused by the cutters may also affect the risk of IRBs. Another factor influencing the difference may be flow through the probes. Maximum flow through the 20 G probes was roughly two times higher than through the 23 G probes and approximately six times higher than 25 G probes [51]. The smaller instruments and slower fluidics in MIVS may reduce the occurrence of IRBs. Additionally, the 20 G cutter typically has a tip to port distance of 0.38 mm, while the corresponding distance for the 23 G cutters is typically 0.23 mm [26]. The reduced cutter tip to port distance allows greater use of cutter to delaminate and dissection epiretinal membranes while cutting, so the scissors are less frequently required during surgery. This difference in surgical technique may reduce the incidence of IRBs as less needed to exchange surgical instruments during surgery. It has to be noted that, in contrast to the incidence of IRBs related to other factors, the rate of retinal breaks related to PVD induction did not differ between MIVS and 20 G vitrectomy in this meta-analysis. The surgical maneuvers for initiating and propagating induction of PVD are essentially the same, independent of the gauge of the vitreous instruments, so the propensity for occurrence of retinal breaks associated with surgically induced PVD may not differ even with a small-incision vitrectomy system [9, 47].

Postoperative RD is the most common vision-threatening complication after vitreoretinal surgery [38]. The rate of postoperative RDs was significantly lower in the 23 G/25 G MIVS group than in the 20 G vitrectomy group in this meta-analysis. The higher rate of postoperative RDs which were associated with 20 G vitrectomy may be accounted for the higher rate of IRBs encountered in this type of surgery, for some of them may later result in RDs. The oblique, tunnelized, smaller size of the sclerotomy in MIVS may reduce the possibility of incarceration of a vitreous wick in the sclerotomy, when compared with a traditional direct, sutured incision. Vitreous incarcerated in the sclerotomy or around the sclerotomy may cause shrinkage and consequently induce a retinal break postoperatively, which can result in an RD [52]. In the subanalysis, we found a decreased risk of postoperative RDs in 23 G MIVS compared to 20 G vitrectomy, but not for 25 G MIVS. The flexibility of the 25 G instruments makes the removal of the peripheral vitreous more difficult than 20 G instruments, while such problem is less pronounced with the 23 G system because of the increased stiffness of the tools. The 23 G MIVS possibly makes a better cleaning of the vitreous around the sclerotomies at the conclusion of vitrectomy than 25 G MIVS, which may also influence the incidence of RDs.

Several limitations of this work should be acknowledged. First, a potential source of heterogeneity is the different type of surgical indications included in this analysis, although we found similar results in the subgroup analysis based on macular surgery or nonmacular surgery, the results should be interpreted with caution. Second, most of studies included in the meta-analysis were retrospective. Retrospective studies have inherent biases and may not accurately reflect the true incidence of IRBs and postoperative RDs. Thus, prospective or randomized clinical trials of larger sample size are needed to provide more definitive information. Third, it was not possible to stratify the studies based on types of vitrectomy cutters used (as the type of cutters are rarely reported). As shown by Teixeira et al. [50], electric 20 G cutters have generally higher range of traction compared to pneumatic 20 G cutters. Therefore, in theory, it is possible that use of electric 20 G cutters accounts for most (if not all) of the increased incidence of IRMs and RDs. An additional limitation is that many studies do not report a detailed list and a breakdown of the vitreoretinal conditions used as an indication for surgery, which limited the current study to a subanalysis of macular vs. nonmacular conditions. Finally, the majority of the studies included in the meta-analysis are based on relatively small numbers. However, there was no obvious tendency for series with fewer cases to show different results compared with reports including more cases.

In conclusion, our findings indicate that MIVS is associated with a much lower incidence of IRBs compared to conventional 20 G vitrectomy for vitreoretinal surgery, especially for sclerotomy-related retinal breaks. Similarity, MIVS has a lower incidence of postoperative RDs, especially for 23 G MIVS.

Summary

What was known before

  • The microincision vitrectomy surgery may carry similar risk of choroidal detachment and vitreous hemorrhage. But iatrogenic retinal breaks and postoperative retinal detachments are uncertained.

What this study adds

  • The microincision vitrectomy surgery may be superior to conventional 20-gauge vitrectomy in reducing the incidence of iatrogenic retinal breaks and postoperative retinal detachments.