Introduction

The pathogenesis of bronchopulmonary dysplasia (BPD) is multifactorial and is characterized by an arrest of the normal growth and development of the immature lung, and lung injury due to oxygen exposure, “ventilator-induced lung injury” (VILI) and prenatal and postnatal inflammation [1]. VILI is multifactorial, and is believed to be due to a combination of volutrauma and atelectotrauma [2]. Furthermore, overdistension of the lung stimulates the release of multiple products that promote an inflammatory response. Subsequently, these proinflammatory cytokines attract peripheral leukocytes into the lung amplifying preexisting lung injury and interfering with alveolarization (biotrauma) [3, 4].

It is well known that VILI may be reduced by the use of gentle ventilation strategies based on reducing the magnitude and duration of mechanical ventilation; one of these ventilation strategies consists of using faster ventilator rate to decrease as much as possible tidal volume and to reduce volutrauma and the release of inflammatory mediators, while achieving adequate gas exchange [5, 6].

High-frequency oscillatory ventilation (HFOV) was developed as a promising ventilation technique to decrease VILI, using tidal volumes smaller than anatomical dead space, delivered at frequencies above normal physiological breathing, and superimposed on a continuous positive pressure [7]. Unlike initial results in animal models [8, 9], the use of elective HFOV compared with conventional mechanical ventilation (CMV) has not succeeded in demonstrating significant differences in decreasing the risk of BPD in preterm infants. This lack of difference has presumably been attributed to the use of different interventions and individual patients across trials [10] and, what more importantly, to the use of devices that did not measure high-frequency tidal volume (VThf) that might have been larger than the anatomical dead space [11].

During HFOV, CO2 removal is directly proportional to the square of the VThf and the frequency (fR) by the following equation: DCO2 = VThf2 × fR. In the last few years, the emergence of volume guarantee (VG) combined with HFOV has allowed independent adjustment of VThf and frequency [12, 13]. In this way, it is possible to modify the frequency without any effect in the VThf; so, pCO2 decreases as the frequency is increased and the VThf is fixed [14,15,16]. Furthermore, it has been possible to develop a new theoretical lung protective strategy based on the use of frequencies higher than normally used (> 15 Hz) allowing to reduce the VThf to theoretically minimize volutrauma while maintaining an adequate ventilation (keeping a constant DCO2 and PCO2) [17]. To maintain the VThf set at higher frequencies, a higher proximal oscillation pressure amplitude (ΔPhf) in the circuit could be needed, but it has been demonstrated that distal pressure amplitude does not increase, it may even decrease [18].

So, with this new strategy, very small VThf can be used in a constant way while the frequency is increased to maintain adequate ventilation, and this very small VThf can be protective compared with the standard VThf generated by using lower frequencies.

The aim of the study was to demonstrate the potential protective effect of this new strategy compared with the standard HFOV, by examining histological changes of acute lung injury in a neonatal animal model after 12 hours of ventilation.

Materials and methods

Study subjects

Twelve healthy 24- to 48-h old Landrace-Large White piglets were used. Animals were handled according to the European and National Regulations for Protection of Experimental Animals (2010/63/UE and RD 53/2013) and the study was approved by the institutional ethic committee for animal research.

Anesthesia

General anesthesia was induced via facemask with 8% sevoflurane in oxygen. Then, anesthesia was maintained with an intravenous constant infusion of propofol as needed. Intraoperative analgesia was achieved with intravenous morphine chlorhydrate (1 mg/kg), and muscle relaxation was induced with intravenous rocuronium (1.5 mg/kg) in this animal model to better analyze lung mechanics without the interference of respiratory muscle activity.

Experimental protocol

When an adequate level of anesthesia was achieved, a tracheostomy was performed and a tracheal tube (3 mm inner diameter) was placed and sealed with a ribbon wrapped around the trachea to prevent any leak. Furthermore, by surgical cut down, a 24-gauge polyethylene catheter was placed into the internal jugular vein for continuous infusion of dextrose 5% (10 mL/kg/h) and the administration of drugs, and a 20-gauge polyethylene catheter was inserted into the right carotid artery, for arterial blood sampling and continuous blood pressure measurement via a calibrated pressure transducer.

Then, the ventilator (Babylog VN500®, Dräger, Lübeck, Germany) was connected to the tracheal tube and adjusted in CMV using Pressure support ventilation (PSV) with VG mode with the following ventilator settings: a positive end-expiratory pressure (PEEP) of 5–6 cmH2O, an inspired oxygen fraction (FiO2) of 0.3–0.4, a frequency of 40–60 rpm, and a tidal volume (VT) of 6–7 mL/kg. Fine adjustments were done for a target arterial partial pressure of carbon dioxide (PaCO2) of 45–60 mm Hg and a target oxygen transcutaneous saturation (SpO2) of 90–95%.

Following instrumentation and after a stabilization period of 60 min, bronchoalveolar lavage (BAL) with three aliquots of 10 mL/kg isotonic saline solution warmed to body temperature was used to induce surfactant depletion mimicking neonatal respiratory distress syndrome. Lung dynamic compliance (Cdyn) was measured before and after BAL to verify a low compliance situation. Then, FiO2 was increased until 1 and the ventilator was switched to HFOV to carry out a lung volume recruitment maneuver in order to ensure adequate alveolar re-expansion. At first, mean airway pressure (mPaw) was increased from 10 to 25 cm H2O in three 1-min stages. Subsequently, FiO2 was decreased until 0.4 and mPaw was decreased too by 2 cm H2O every 30 s until 10 cm H2O or 2 cm H2O above the critical lung closing pressure (defined as the pressure below which SpO2 decrease below 90%).

Finally, after lung recruitment was achieved, ventilator settings were adjusted according to the ventilation strategy assigned using random numbers to each group. Three groups were used to better analyze the effect of the two strategies of HFOV and the standard CMV in a similar lung damage situation as a reference.

  1. 1.

    CMV group: Animals assigned to this group were ventilated in CMV using PSV with VG mode with the same ventilator settings described above adjusted for the target PaCO2 of 45–60 mm Hg.

  2. 2.

    HFOV groups (10Hz and 20Hz HFOV groups): Animals assigned to other two groups were ventilated in HFOV with VG and adjusted to a mPaw of 10 cm H2O, an inspired oxygen fraction of 0.3–0.4, an IE ratio of 1:1, a frequency of 10 or 20 Hz, and the VThf guarantee needed to achieved the target PaCO2 (45–60 mm Hg). Delta pressure (∆Phf) was adjusted by the ventilator to achieve the set VThf.

A 20-gauge polyethylene catheter was advanced to the carina, between the tracheal tube and the trachea for continuous measurement of mPaw at the trachea (mPaw-t) and ∆Phf at the trachea (∆Phf-t). Self-adhering patches were applied to the skin for electrocardiography (ECG) and heart rate recording. Pulse oximetry was recorded continuously by placing a pulse oximeter on the paw, and rectal temperature was also monitored and maintained between 36 °C and 38 °C by means of a total temperature management system. Temperature, pulse oximetry, heart rate, arterial blood pressure and airway pressure (referred to a zero level) were registered continuously on a cardiovascular monitor (PM8060 Vitara, Dräger, Lübeck, Germany). Arterial blood was withdrawn anaerobically and immediately analyzed for pH, partial pressure of oxygen (PaO2), PaCO2, and arterial oxyhemoglobin saturation (SaO2) each 2 h, ensuring an adequate ventilation and oxygenation (IL 1306 pH/Blood GasAnalyzer, Allied Instrumentation). Data from the ventilator were exported through the standard USB connection.

After 12 h of ventilation according to the strategy assigned to each group, a new Cdyn measurement was repeated under the same conventional settings on the ventilator. Then, animals were euthanized using a propofol overdose followed by an intravenous bolus of potassium chloride solution (50 mg/kg).

The HFOV was delivered with a Babylog VN500 ventilator (Dräger, Lübeck, Germany). The Babylog VN500 generates a sinusoidal pressure signal around a set mPaw and has an active inspiration and an active expiration. The VG mode is volume-targeted ventilation where the microprocessor compares the VThf of the previous breath, using leak compensated VThf, and adjusts delta pressure up or down to achieve the set VThf.

Tissue removal and lung processing

After midline sternotomy, lungs were removed and inflated with saline solution at a pressure of 20 cm H2O. Subsequently, lungs were perfused with saline solution and then, the right lung was fixed with formaldehyde 4%. Small pieces from medial lobe of this lung were removed and stored in the fixation solution at −4 °C. Then, the specimens were embedded in paraffin, sectioned, and stained with hematoxylin and eosin for histological examination.

Histological examination

Lung histological damage was quantified by a score using five variables (inflammation, edema, hemorrhage, atelectasis, and emphysema). A predefined injury severity was created and graded for each of the variables [19] as shown in Table 1 (total score between 0 and 20). The pathologists were blinded to the assignment group.

Table 1 Defined lung histological examination score

Statistical analysis

Initially, a sample size of six animals per group was considered adequate based on previous studies. An interim analysis was done by the research team once were included four animals per group to prevent extra experiments; as statistically significant differences were observed at this point, only four animals were used per group, in order to reduce the number of animals used according to the European and National Regulations for Protection of Experimental Animals.

Statistical analysis of data was performed using SPSS 15.0 (SPSS Inc., IL, USA). Following confirmation of normal distribution (Shapiro–Wilk test), dynamic compliance was analyzed using a two-way analysis of variance (ANOVA) for repeated measures followed by Tukey’s test for multiple comparisons. The mean VThf in HFOV groups was analyzed using a t-test for unpaired data. These data are presented as means ± standard deviation. For non-parametric data (lung histological score comparison), an ANOVA of Kruskall–Wallis was used. Statistical significance was accepted when p < 0.05.

Results

The mean (± standard deviation, SD) body weight of the piglets were 2.57 ± 0.59 kg. All animals were adequately ventilated and oxygenated. There were no significant differences in hemodynamic parameters between groups.

Cdyn was measured pre- and post-BAL, as well as at the end of the study. As shown in Fig. 1, BAL significantly reduced Cdyn in a similar way in all the groups. At the end of the experiment, Cdyn improved in all groups.

Fig. 1
figure 1

Dynamic respiratory compliance (Cdyn) pre-BAL, post-BAL, and final. Cdyn was significantly lower in post-BAL compared with pre-BAL for CMV (2.29 ml/cm H2O vs 1.19 ml/cm H2O; p = 0.001*), 10 Hz (2.81 ml/cm H2O vs 1.24 ml/cm H2O; p = 0.035*), and 20 Hz groups (2.19 ml/cm H2O vs 0.75 ml/cm H2O; p = 0.006*). BAL bronchoalveolar lavage, CMV conventional mechanical ventilation, HFOV high-frequency oscillatory ventilation. *p  < 0.05

CMV group

Piglets in the CMV group needed a VT of 7.33 ± 0.20 ml/kg to achieve the target PaCO2 of 45–60 mm Hg.

Emphysema was the main lesion in this group. Inflammatory infiltrate and atelectasis were also found in this group. Neither hemorrhage nor edema were found in the animals of this group (Table 2 and Fig. 2).

Table 2 Lung histological examination of HFOV and CMV groups
Fig. 2
figure 2

Lung histological examination. a, b 10 Hz high-frequency oscillatory ventilation (HFOV) group. Severe diffuse inflammatory infiltrate and hemorrhage. c, d 20 Hz HFOV group. Medium atelectasis. e, f Conventional mechanical ventilation (CMV) group. Moderate emphysema with large alveolar rupture

HFOV groups

Piglets in the 20 Hz group needed a lower VThf than those in the 10 Hz group to achieve the target PaCO2 of 45–60 mm Hg (1.56 ml/kg vs 2.54 ml/kg; p = 0.004).

More evident and more severe histological lesions were found in the 10 Hz group with the higher VThf. Inflammatory infiltrate was the main lesion in this group, including mononuclear infiltrate of lymphocytes, plasma cells, and macrophages within the alveolar wall and alveolar space, as well as large areas of parenchyma consolidation. Areas of alveolar hemorrhage were found in the more severe cases. Finally, areas of atelectasis were found in most cases. The more severe the inflammation and hemorrhage, the more severe atelectasis was found. Edema and emphysema were not found in this group.

The number of samples with inflammatory infiltrate, as well as its severity in the 20 Hz groups was less than in the 10 Hz group. Atelectasis was the main lesion in the 20 Hz group with the lower VThf, but mild or medium. No hemorrhage, edema, and emphysema were found in this group (Table 2 and Fig. 2).

When lung histological score was compared between groups, statistically significant differences were observed for inflammation (p = 0.043), hemorrhage (p = 0.027), and emphysema (p = 0.027).

Discussion

We have demonstrated a reduced lung injury histological score with the use of a strategy of lower compared with higher VThf on HFOV combined with VG in an animal neonatal model of lung injury during mechanical ventilation. It has become increasingly evident that lung overdistension by the use of large gas volume is an important factor to induce VILI [20]. Furthermore, Wakabayashi et al. demonstrated that lung-marginated, intravascular monocytes are only activated following lung overstretch, not atelectasis, leading to proinflammatory cytokines production that promote VILI and nonpulmonary organs injury (“biotrauma”) [21]. So, the use of strategies to reduce as much as possible the tidal volume may minimize lung and systemic injury.

In contrast to HFOV, emphysema was the main lesion in CMV group; inflammatory infiltrate and atelectasis were also found in this group, but milder. Previous experimental models that compared both, HFOV without VG at 10 Hz [22] and 15 Hz [23] and CMV using a protection ventilator strategy with low tidal volume (5–6 mL/kg) and high PEEP (8–10 cm H2O) have shown that the use of HFOV was associated with an attenuation of lung inflammation (polymorphonuclear infiltration and tumor necrosis factor-α production in the alveolar space), histopathological changes of the small airways and alveoli and oxidative lung injury. These authors suggested that a more homogeneous lung inflation due to an adequate lung volume recruitment in HFOV may limit alveolar stretch minimizing VILI. No significant differences of inflammation were found between 20 Hz HFOV and CMV groups in our model; however, no emphysema was found in 20 Hz HFOV group compared with the CMV group and the degree of lung histological damage was greater in CMV group based on our developed score (20 points vs 11 points). We hypothesize that the lower inflation pressure at the alveoli, the minor variations in pressure and volume and the relatively constant lung volume above the functional residual capacity achieved in HFOV compared with CMV, may explain these differences.

We found that when a higher VThf was used in HFOV, there was more evident and severe inflammatory infiltrate, including mononuclear infiltrate of lymphocytes, plasma cells, and macrophages within the alveolar wall and alveolar space, as well as large areas of parenchyma consolidation. Meyer et al. compared lung lesions resulting from two different HFOV regimes in a rabbit model, at 5 Hz and 15 Hz without the combined use of VG, so the higher the frequency, the lower the VThf. They observed a significantly lower neutrophil count within alveoli in animals ventilated with the higher frequency regime (and lower VThf) after 4 h [24]. Furthermore, in a large animal model of acute respiratory distress syndrome, HFOV at higher frequencies (9 Hz vs 3 and 6 Hz) delivered smaller tidal volumes and minimized stress on the lung by reducing lung injury and the expression of inflammatory mediators. Lung histopathology revealed less hemorrhage and neutrophil infiltration in the alveoli and interstitium and less interstitial lymphocyte infiltration in animals ventilated under HFOV at 9 Hz and smaller tidal volumes, compared with lower frequencies, as we showed in our model [25].

So, based on our results, the use of very low tidal volume in HFOV combined with VG, together with an adequate lung volume recruitment, may reduce lung damage in immature infants.

The main limitation of our model was the short duration of the ventilator strategy used, 12 h, which only shows the acute lung injury. However, it is well known that inflammation occurs shortly after birth, driven by antenatal factors in combination with early postnatal exposure to intensive care interventions such us mechanical ventilation; and that this early inflammation plays a critical role in the development of BPD [26]. Thus, any intervention used to reduce lung inflammation in this vulnerable period may be effective to reduce lung damage and BPD. Furthermore, the neonatal animal model used in this study is not defined to reproduce BPD, but it is useful to analyze acute lung damage of mechanical ventilation.

In conclusion, the use of very low tidal volume with increasing frequencies in HFOV combined with VG reduces inflammatory response and lung histological damage. So, this new protective lung strategy may allow minimizing VILI in immature infants in the immediate neonatal period.