Thank you for visiting nature.com. You are using a browser version with limited support for CSS. To obtain the best experience, we recommend you use a more up to date browser (or turn off compatibility mode in Internet Explorer). In the meantime, to ensure continued support, we are displaying the site without styles and JavaScript.

  • Article
  • Published:

Validation of the PHYSIO-PORT UP ambulatory blood pressure monitor in pregnant women according to the European Society of Hypertension International Protocol revision 2010

Abstract

The present study examines the performance of the ambulatory blood pressure monitor PHYSIO-PORT UP in pregnant and preeclamptic women. The systolic blood pressure (SBP) and the diastolic blood pressure (DBP) of 33 subjects (among them 13 with preeclampsia) were sequentially measured and compared to reference measurements. Those reference measurements were taken by two observers using a mercury sphygmomanometer. Subject recruitment and evaluation of the measurement results were performed according to the European Society of Hypertension International Protocol revision 2010 (ESH-IP 2010). Three readings were recorded per subject under static conditions. The protocol did not include testing under ambulatory conditions. 87/90 (SBP/DBP) of the 99 readings had a difference ≤ 5 mmHg, 96/98 had a difference 98 ≤ 10 mmHg, and 98/99 had a difference ≤ 15 mmHg. 2 out of 3 readings ≤ 5 mmHg was achieved by 31/32 subjects. None of the subjects had less than 1 comparative reading ≤ 5 mmHg. In conclusion, the PHYSIO-PORT UP fulfilled all requirements of the ESH-IP 2010. According to these results, the PHYSIO-PORT UP can be recommended for blood pressure measurement in pregnant women.

This is a preview of subscription content, access via your institution

Access options

Buy this article

Prices may be subject to local taxes which are calculated during checkout

Fig. 1

Similar content being viewed by others

References

  1. O’Brien E, Atkins N, Stergiou G, Karpettas N, Parati G, Asmar R, et al. Working Group on Blood Pressure Monitoring of the European Society of Hypertension. European Society of Hypertension International Protocol revision 2010 for the validation of blood pressure measuring devices in adults. Blood Press Monit. 2010;15:23–38.

    Article  Google Scholar 

  2. Abou-Dakn M, Döhmen C, Wenzel S. Validation of the TONOPORT VI ambulatory blood pressure monitor in adults according to the European Society of Hypertension International Protocol revision 2010. J Hum Hypertens. 2017;31:89–92.

    Article  CAS  Google Scholar 

Download references

Acknowledgements

PAR Medizintechnik GmbH & Co. KG funded this study and supplied a PHYSIO-PORT UP device free of charge for the purpose of this study.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Michael Abou-Dakn.

Ethics declarations

Conflict of interest

The authors declare that they have no conflict of interest.

Electronic supplementary material

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Abou-Dakn, M., Wenzel, S. Validation of the PHYSIO-PORT UP ambulatory blood pressure monitor in pregnant women according to the European Society of Hypertension International Protocol revision 2010. J Hum Hypertens 32, 770–774 (2018). https://doi.org/10.1038/s41371-018-0101-y

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1038/s41371-018-0101-y

This article is cited by

Search

Quick links