Abstract
Utilizing a pinhole-plane imaging concept, Airyscan allows for simultaneous improvement in resolution and signal-to-noise by capitalizing on an innovative 32-channel GaAsP photomultiplier tube (PMT) array detector. Each detection channel functions as a very small pinhole to increase resolution while the overall detector design delivers better signal-to-noise than traditional GaAsP-based confocal systems. In the past, a stack of at least five z-slices had to be deconvolved to get usable images with an optical section thinner than 1 Airy unit. Now, the new 2D Superresolution mode for Airyscan delivers images with the thinnest optical section (0.2 Airy units) from a single image while maintaining the light collection efficiency of a much larger 1.25-Airy-unit pinhole.
Main
The creation of optical section images in biomedical research started in 1955, when Marvin Minsky laid the foundation for confocal microscopy1. In an effort to improve the contrast of images for the study of the central nervous system, Minsky placed an aperture in a conjugate plane relative to the microscope objective. He discovered that he could create an optical section within a three-dimensional biological sample by blocking any light that did not originate from the focal plane. Hence, Minsky was able to generate high-contrast images to study nervous system circuitry (prepared via the Golgi method) free from out-of-focus scatter from imaging planes deep within scattering samples. This confocal concept was the foundation of modern laser scanning confocal microscopy.
Creation of the optical section
Conventional confocal laser scanning microscopes leverage a combination of a physical aperture (the so-called pinhole) and a unitary detector (PMT, hybrid detector, avalanche photodiode, etc.) to create an optically sectioned image2. Naturally, by varying the aperture size, one can also change the thickness of the optical section (Fig. 1a). In addition to the thickness of the optical section, the resulting image resolution and image signal-to-noise are directly related to the aperture size (Fig. 1b,c). As the pinhole is closed, less signal can reach the unitary detector, which significantly reduces the image signal-to-noise. Traditional confocal microscopes demand the sacrifice of optical sectioning and resolution for the sake of a usable signal-to-noise ratio and acceptable sample bleaching and phototoxicity. Consequently, a pinhole size of 1 Airy unit (AU) has become the default for confocal microscopes.
In contrast to traditional confocal detectors, Airyscan combines the resolution benefits of imaging with a small pinhole with the collection efficiency of a large pinhole. This design delivers a simultaneous increase in resolution and signal-to-noise by projecting 1.25 AU onto a 32-channel GaAsP PMT array detector. Each detector element behaves as a 0.2-AU pinhole while the collection efficiency of a 1.25-AU pinhole is maintained3,4 (Fig. 1d). Moreover, linear deconvolution is used to extend the resolution beyond what a 0.2-AU pinhole provides, resulting in a resolution increase in all three spatial dimensions. The new 2D Superresolution mode for Airyscan improves the optical sectioning by a factor of 2.4 for a single optical section by utilizing a new exclusive processing algorithm in place of the linear deconvolution step5,6 (Fig. 1a). Along with improved optical sectioning, Airyscan now provides lateral resolution to 120 nm for 2D and 3D data sets (z-stacks) and 350-nm axial resolution for z-stacks (Fig. 2).
The new 2D Superresolution mode
Since its launch, Airyscan's purpose has been to ensure that all precious fluorescence emission photons are preserved to provide the highest sensitivity of any confocal. This focus on system sensitivity has resulted in unprecedented image quality, resolution and sectioning for 3D stacks. Airyscan 2D images have always offered an improvement in lateral resolution while maintaining an optical slice thickness of 1.25 AU (Figs. 3 and 4).
The new 2D Superresolution mode for Airyscan takes advantage of the fact that the confocal point spread function entangles information from the x, y and z planes in the pinhole plane. Airyscan measures the emission fluorescence distribution already in a single image acquisition, thereby yielding information about how the signal is entangled. The new exclusive algorithm of the 2D Superresolution mode makes it possible to distinguish and separate the light originating in the focal plane from light originating outside of the focal plane. This allows researchers to obtain perfect optical sections, without the need to acquire a z-stack (Figs. 3,4,5). To get the same optical section on a traditional confocal, a researcher would have to close the pinhole to 0.2 AU and sacrifice >95% of all emission signal (in both the plane of interest and out-of-focus planes). This would limit the information that could be obtained from the plane of interest.
Summary
The Airyscan detector from ZEISS delivers more information from the pinhole plane than traditional confocal microscopes. With this information, researchers get simultaneous improvements in resolution and signal-to-noise and an increase in acquisition speeds. The new 2D Superresolution mode improves the achievable resolution of a confocal microscope to 120 nm laterally and delivers a 2.4× thinner optical section. In image stacks, an axial resolution of 350 nm is achieved.
This article was submitted to Nature Methods by a commercial organization and has not been peer reviewed. Nature Methods takes no responsibility for the accuracy or otherwise of the information provided.
References
Minsky, M. Scanning 10, 128–138 (1988).
Conchello, J.-A. & Lichtman, J.W. Nat. Methods 2, 920–931 (2005).
Huff, J. Nat. Methods 12, i–ii (2015).
Huff, J. Nat. Methods 12, i–ii (2016).
Wilhelm, S. Confocal Laser Scanning Microscopy (ZEISS, 2010).
Pawley, J. Handbook of Biological Confocal Microscopy 3rd ed. (Springer, 2006).
Author information
Authors and Affiliations
Corresponding author
Additional information
Disclaimer
This article was submitted to Nature Methods by a commercial organization and has not been peer reviewed. Nature Methods takes no responsibility for the accuracy or otherwise of the information provided.
Rights and permissions
About this article
Cite this article
Huff, J., Bergter, A., Birkenbeil, J. et al. The new 2D Superresolution mode for ZEISS Airyscan. Nat Methods 14, 1223 (2017). https://doi.org/10.1038/nmeth.f.404
Published:
Issue Date:
DOI: https://doi.org/10.1038/nmeth.f.404
This article is cited by
-
Macroscale structural changes of thylakoid architecture during high light acclimation in Chlamydomonas reinhardtii
Photosynthesis Research (2024)
-
SHI family transcription factors regulate an interspecific barrier
Nature Plants (2023)
-
Integrated multimodality microscope for accurate and efficient target-guided cryo-lamellae preparation
Nature Methods (2023)
-
ALS/FTD-associated mutation in cyclin F inhibits ER-Golgi trafficking, inducing ER stress, ERAD and Golgi fragmentation
Scientific Reports (2023)
-
A specific hybridisation internalisation probe (SHIP) enables precise live-cell and super-resolution imaging of internalized cargo
Scientific Reports (2022)