Abstract
Objective:
Electrical cardiometry (EC) is an impedance-based monitor that provides noninvasive, real-time hemodynamic assessment. However, the reference values for neonates have not been established.
Study Design:
EC (Aesculon) was applied to hemodynamically stable preterm and term infants. Hemodynamic variables included cardiac output (CO), cardiac index (CI), stroke volume (SV) and heart rate (HR). Their gestational age (GA), weight and body surface area (BSA) were recorded.
Results:
A total of 280 neonates were studied. Their GA ranged from 265/7 to 414/7 weeks, weight 800 to 4420 g and BSA 0.07 to 0.26 m2. CO was positively correlated to GA, weight and BSA (r=0.681, 0.822, 0.830, respectively; all P<0.001). Using regression analysis, CO was most significantly correlated to BSA. Mean CI was 2.55±0.37 l min−1 per m2.
Conclusion:
Hemodynamic reference by EC is notably distinct among neonates of diverse maturity. CO is most closely correlated to BSA.
Similar content being viewed by others
Introduction
In the care of critically ill patients, hemodynamic assessment is crucial for both diagnosis and management; however, clinical assessment of cardiac output (CO) by the interpretation of indirect parameters such as blood pressure may be misleading. In addition, trend monitoring instead of spot measurements may be more informative.1 The thermodilutional technique has long been recognized as a standard CO measurement in pediatric patients, but the invasiveness and technical impracticability restrict its usefulness in neonates. Noninvasively, functional MRI has shown promise as a tool to accurately access CO, but does not allow for monitoring for prolonged periods. More commonly, noninvasive assessment of CO is estimated by Doppler measurements on echocardiogram, but this is technically demanding and can only be obtained intermittently. Ideally, hemodynamic monitoring in neonates should be accurate, practical, noninvasive, continuous in absolute measurements and feasible to entire spectrum of age and weight.2
Electrical cardiometry (EC) has been proposed as a safe, accurate and reproducible technique for hemodynamic measurement in children and infants.3 It is an impedance-based monitoring device that provides real-time cardiovascular assessment. A high-frequency, low-magnitude, alternative current is generated by two distant electrodes (from head to left lower extremity), whereas two in-between electrodes (positioned at the neck and axilla) measure electrical changes dependent on alterations in thoracic electrical bioimpedance during the cardiac cycle. The changes in electrical bioimpedance is related to aortic flow pattern, and more specifically, influenced by the alignment of red blood cells in the aorta. When aortic flow stops and aortic valve closes, red blood cells are randomly orientated and interfere with electrical conduction. As the left ventricle contracts and aortic valve opens, the ejection of blood forces red blood cells to align in parallel with the flow, and the electrical current in the aorta passes with less impedance, which results in decreased impedance and higher conductivity. The pulsatile impedance waveform corresponds to the cardiac cycle. Moreover, the rate of impedance change is used in the calculation of hemodynamic measures, such as blood velocity, contractility, SV and CO.4, 5
Clinical applications of EC have been widely studied. In comparison with the thermodilution technique, the usefulness of EC in providing hemodynamic measurements has been verified in piglet models6 and adult population.7 Furthermore, a good correlation was found between CO measurements by EC and pulmonary artery catheter thermodilution in children receiving catheterization.8 EC has been shown to be comparable to direct Fick-oxygen method9 and transesophageal Doppler10 in CO measurements in infants with congenital heart disease. By using transthoracic echocardiogram, the accuracy of EC in children,11, 12 neonates13, 14 and even preterm infants15, 16, 17 has been demonstrated. A recent study indicates that SV measurements by EC significantly correlated with measurements by transthoracic echocardiogram in preterm infants with gestational age (GA) 25 to 34 weeks.18 Previously, we used EC to continuously monitor preterm infants during surgical ligation of the patent ductus arteriosus and found a significant CO deterioration upon termination of ductal shunting.19 Cote et al.20 also demonstrated clinical usefulness of EC in monitoring hemodynamics for children undergoing surgery.20 Although the utilization of EC in the care of critically ill patients have gained popularity, normal hemodynamic parameters based on EC for neonates of different age and weight have not been reported. The clinical utilization relies heavily on measurement trends and not absolute values. Thus, we conducted this observational study to determine the reference ranges for hemodynamically stable neonates of diverse maturity and weight.
Methods
Study population
This prospective study was conducted in the neonatal department of Chang Gung Children Hospital from January to July 2013. The neonatal unit is a referral center with 37 intensive care beds, 70 sick-baby nursery beds and an annual admission of ~1800 infants. Our candidates were newborn neonates, either inborn or outborn, who were admitted before 72nd hour of life, and were evaluated for cardiovascular status between postnatal age third to fourth day of life. Only neonates with normal blood pressure according to Philadelphia Neonatal Blood Pressure Study Group,21 normal heart rate, normothermia and adequate urine output >1 ml kg−1 per day were considered hemodynamically stable and enrolled into the study. We excluded those with major malformations, inotrope support or sedative, structural heart disease other than a patent foramen ovale or atrial septal defect, excessive weight loss defined as >10% of birthweight, Apgar score <7 at 5-min, clinical evidence of perinatal asphyxia, ultrasound-confirmed intraventricular hemorrhage, bacteremia, hydrops fetalis or pleural effusion. Neonates with intrauterine growth restriction (IUGR, i.e., neonates born with birthweight <10th percentile for GA) were also identified. Institutional Review Board approved this study and written informed consent was obtained.
To evaluate the impact of invasive or noninvasive ventilator support on hemodynamic status, we performed a preliminary comparison before the design of the study. Adjusting for distinct demographic characteristics, we recognized that neonates with invasive mechanical ventilation had lower mean arterial pressure (MAP) and CO. However, there was no statistically hemodynamic difference between neonates breathing room air, using nasal cannula or with noninvasive ventilation support (i.e., continuous positive airway pressure, CPAP). Consequently, we also excluded neonates with invasive respiratory support in our study.
EC setting
EC (Aesculon; Osypka Medical, Berlin, Germany) was applied by means of four standard surface electrocardiogram electrodes over the infants’ forehead, left lower neck, left mid-axillary line at the level of xiphoid process and lateral aspect of left thigh. EC was set to record measurements at 1-min intervals. Sixty cardiac cycles were used for averaging parameters and for stroke volume variation (SVV) calculation. Default central venous pressure (CVP) by Aesculon was 3 mm Hg. This value was used in the calculation of systemic vascular resistance (SVR), based on the following formula: SVR=80 x (MAP−CVP)/CO, reported in units dyn·s cm−5.
Parameters
Demographic descriptions, including gender, GA, weight, body surface area (BSA) and Apgar scores, were recorded. Weight at the time of experiment or birthweight, whichever was heavier, was adopted for EC calculation. BSA was computed according to the Boyd formula by Aesculon. Their respiratory condition was labeled as breathing room air, nasal cannula or nasal CPAP.
Hemodynamic parameters comprised of CO, cardiac index (CI), heart rate (HR), SV, SVV, thoracic fluid content (TFC), index of contractility (ICON) and SVR were provided by EC. CI is the product that relates CO to BSA (CO/BSA) and is used for evaluating cardiac performance among infants of different size. TFC is derived from the thoracic electrical base impedance (1/base impedance), which is dependent on thoracic intravascular and extravascular fluid. Larger TFC indicates a higher total thoracic fluid volume. ICON is derived from the maximum rate of change of thoracic electrical impedance. As ventricular contractility increases, the impedance falls at a faster rate, which results in a higher ICON value. Both TFC and ICON are unitless. Each individual’s concurrent MAP and axillary temperature were also gathered.
Measurement protocols
We performed EC measurement for each study neonates between postnatal age 72 to 96 h. The timing of examination was chosen to reduce the hemodynamic impact of patent ductus arteriosus, as most physiologic ductus arteriosus were functionally closed before age 24 to 48 h.22 For neonates weighting <1500 g or <32 weeks of gestation, echocardiogram were performed to confirm closure of the ductus. We executed the measurements in the supine position during sleep, if feasible, or consoled them to avoid agitation. Phototherapy, if applied, was discontinued temporarily during the measurement. Blood pressure was recorded noninvasively with appropriate neonatal cuff immediately before EC measurement.
We defined at least five continuous and qualified signals (i.e., signal quality index, provided by Aesculon, ⩾70%) before starting measurement as steady condition. Once one’s condition was steady, five continuous and qualified measurements over the duration of 5 min were collected and subsequently averaged.
Data analysis and statistics
Study neonates were sorted by GA (⩽28, 29 to 30, 31 to 32, 33 to 34, 35 to 36, 37 to 38, 39 to 41 weeks), weight (<1000, 1000 to 1499, 1500 to 1999, 2000 to 2499, 2500 to 2999, 3000 to 3499, ⩾3500 g) and BSA (⩽0.1, 0.11 to 0.13, 0.14 to 0.16, 0.17 to 0.19, 0.20 to 0.22, ⩾0.23 m2).
Statistical analysis was performed using IBM SPSS Statistics version 20 (Armonk, NY, USA). Correlation between parameters was assessed by Pearson's coefficient. When determining the relation between CO and GA, weight and BSA, and between CO and HR and SV, multiple regression analysis using step-wise model was also applied to adjust correlation. Statistical significance was defined as P<0.05.
Result
A total of 280 hemodynamically stable neonates (148 male and 132 female; 118 term and 162 preterm babies) were recruited into the study. Their median (range) GA, weight and BSA were 36 (26 to 41) weeks, 2370 (800 to 4420) g and 0.17 (0.07 to 0.26) m2, respectively. Their Apgar scores (mean±s.d.) were 8±1 and 10±1 at 1 and 5 min of life. There were 29 IUGR neonates in this study. The demographic distribution and respiratory status are summarized in Table 1.
Hemodynamic measurements grouped by GA, weight and BSA are listed in Tables 2, 3 and 4, respectively.
CO, CI and their correlation to GA, weight and BSA
CO ranged from 0.14 l min−1 (GA 27 weeks, 800 g, BSA 0.07 m2 preterm baby) to 0.94 l min−1 (GA 406/7 weeks, 4410 g, BSA 0.25 m2 infant). The CO reference for hemodynamically stable neonates was diverse. CO was significantly correlated to GA, weight and BSA (r=0.681, 0.822, 0.830, respectively; all P<0.001), which verified that larger neonates generally have higher CO. After regression analysis, CO was best correlated to BSA (adjusted r2=0.688, P<0.001; Figure 1).
CI was not associated with GA, weight or BSA, and mean CI was 2.55±0.37 l min−1 per m2 of our neonates. Moreover, we applied CI to compare cardiac performance between IUGR and non-IUGR neonates, and found that there was no significant difference in CI between the two groups (Table 5).
SV and HR
SV and HR ranged from 0.94 to 6.43 ml and 101 to 177 beats per minute, respectively. Testing for correlation, SV correlated with GA, weight and BSA (r=0.488, 0.558, 0.563, respectively; all P<0.001), which further confirms the relationship between SV and the individual’s body size, whereas HR had a negative correlation with GA, weight and BSA (r=−0.441, −0.453, −0.441, respectively; all P<0.001). Importantly, although HR decreased with GA and weight in our study, CO increased due to an increase in SV.
However, SVV varied greatly in our neonates; it ranged from 5.3 to 27, with a mean of 15.8±4.4. There was no significant correlation between SVV and GA, weight, BSA, CO, CI or SV.
Thoracic fluid content
TFC ranged from 15.1 of our smallest neonate to 41.4 of a term infant. TFC had a poor correlation with GA, weight and BSA (r=0.264, 0.356 and 0.369, all P<0.001), and moderate correlation to CO and CI (r=0.511 and 0.415, P<0.001).
Index of contractility
The value of ICON varied between individuals, and it had mild negative correlation to GA, weight or BSA (r=−0.332, −0.345 and −0.345, all P<0.001). It had no association to CO; however, it was moderately correlated to CI (r=0.678, P<0.001).
Systemic vascular resistance
SVR had a negative correlation with GA, weight and BSA (r=−0.398, −0.527 and −0.545; all P<0.001), and also a negative correlation with CO and CI (r=−0.749 and −0.618; P<0.001).
Discussion
CO drives global oxygen delivery and a lower than reference value could serve as a negative prognostic indicator, which is highly valuable in the intensive care setting.23 We recognize that the absolute value of CO measured by bioimpedance, echocardiogram, thermodilution or Fick’s method are not interchangeable24 and parameters may vary between different measuring methods.12 Establishing reference parameters for EC is extremely important for clinicians, as measurements that fall out of the said reference range may be alarming.
We also believe that the reference value should be maturity-specific or body weight-specific. Unlike adults or older children, body size or maturity differs greatly from an extremely preterm baby to a post-term infant. This influences cardiac function accordingly. Neonates are therefore a special population that deserves a precise and comprehensive hemodynamic reference. Our study is the first to exhibit hemodynamic measurements for neonates of different size and age by the bioimpedance method.
In consideration of the hemodynamic influence from different modalities of noninvasive respiratory support, we found that there was no significant CO dissimilarity among neonates with ambient breathing or CPAP. This is in line with previous echocardiographic studies where circulation compromise was not demonstrated in preterm infants using CPAP.25 In contrast, invasive mechanical ventilation, positive end expiratory pressure can affect cardiac function and cause a fall in CO and SV,26 but the degree of ventilation-associated hemodynamic compromise warrants further and more comprehensive hemodynamic investigation.
In our study, we demonstrated that the reference for neonatal hemodynamic measurements should be assessed according to individual age and body size. CO, SV and TFC were positively correlated to GA, weight and BSA, whereas HR, ICON and SVR were negatively correlated. In larger neonates, HR was generally lower and a relatively higher SV contributed to a higher CO. Because vascular resistance depends on vessels’ diameter, it is inevitable that smaller neonates will have greater SVR and larger neonates have lesser SVR. Notably, we conclude that grouping CO measures based on BSA is more suitable than GA or body weight. Therefore, to compare cardiac performance between neonates, we recommend the use of CI, an indicator related to BSA, to minimize maturity difference. Furthermore, we found that the use of CI as a reference applies to IUGR infants as well, as no significant difference in CI was found between IUGR and non-IUGR infants. However, the utility of EC to evaluate specific contributing factors of CO, that is, preload, contractility and afterload, should be discussed.
Preload or intravascular volume is not assessed by EC. Preload is the end volumetric pressure that stretches the ventricles, which is usually measured as end diastolic volume in echocardiogram. There is no similar measurement to be achieved by EC. Although EC provides TFC as an indicator of total thoracic fluid, its usefulness is to evaluate either pulmonary fluid overload (pulmonary edema) or soft tissue edema. Larger neonates have a higher TFC; however, it does not adequately represent intravascular volume and the relationship between total thoracic fluid and intravascular volume is not clear. Another indicator, SVV, has been proposed as a predictor of fluid responsiveness for ventilated patients under surgery:27 the greater the variation, the better the response to fluid expansion, suggesting a low preload status. In our experience, SVV fluctuates quickly when the neonate is active or crying, making it difficult to interpret.
EC only gives theoretical information about cardiac contractility, represented by the parameter ICON, based on the rate of change of thoracic impedance. Myocardial contractility denotes the ability of ventricle to contract and pump, which is commonly represented as ejection fraction in echocardiogram. Although ICON and CI are positively correlated in our study, its correlation to ejection fraction in echocardiogram is still unknown. EC also provides indirect measures of LV performance by pre-ejection period, left ventricular ejection time and systolic time ratio (by means of pre-ejection period/left ventricular ejection time),28 but HR will notably affected by these parameters. When HR increases, pre-ejection period and left ventricular ejection time shorten, and systolic time ratio changes accordingly. As reference range of HR varied among cases, pre-ejection period, left ventricular ejection time and systolic time ratio were unavoidable affected as well, which limit their usefulness in comparison. Therefore, we remain speculative in the usefulness of these three parameters by EC in the neonatal population.
Vascular resistance is one of the factors that determine afterload, and SVR by EC is calculated by the formula: 80 x (MAP−CVP)/CO, under the assumption that CVP is 3 mm Hg. We recognize that such assumption may lead to imprecise SVR calculations. However, SVR can still be useful in trending changes within individual patients. Several limitations should be addressed. First, we did not compare EC with other conventional methods of measurements such as echocardiogram. As plenty of studies have tested the same device in pediatric population,8, 9, 10, 11, 12, 13, 14, 15, 16, 17, 18 we did not intend to verify its effectiveness in our study. Furthermore, our results were compatible with previous comparison studies. Noori et al.13 had performed paired measurements in 20 healthy neonates and concluded that EC is as accurate in measuring CO as echocardiogram. In that study, the mean CO was 0.534 l min−1 for term infants weighing ~3 kg. Our results were comparable: for neonates aged GA ⩾39 weeks, mean CO is 0.49±0.1 l min−1 and for those weighed 3000–3499 g, mean CO is 0.53±0.14 l min−1. However, we acknowledged that more data are needed to demonstrate accuracy, precision and utility of EC before it can be applied in clinical settings. Second, we excluded neonates who were invasively ventilated. As mentioned previously, insufficient understanding of ventilator-associated influence on EC may limit it usefulness in some critically ill neonates. Third, the reference could not be fully applied in neonates with structural heart disease other than a PFO or ASD. Because hemodynamics may be significantly altered among those with structural heart disease, measurements by EC should be interpreted with caution.
In conclusion, hemodynamic reference is notably distinct among neonates of differing maturity. We established the hemodynamic reference for EC, and found CO to be closely correlated to BSA. The reference table can aid in the bedside evaluation of neonatal cardiac performance and in clinical decision-making.
References
de Boode WP . Clinical monitoring of systemic hemodynamics in critically ill newborns. Early Hum Dev 2010; 86 (3): 137–141.
Soleymani S, Borzage M, Noori S, Seri I . Neonatal hemodynamics: monitoring, data acquisition and analysis. Expert Rev Med Devices 2012; 9 (5): 501–511.
de Boode WP . Cardiac output monitoring in newborns. Early Hum Dev 2010; 86 (3): 143–148.
Osypka MJ, Bernstein DP . Electrophysiologic principles and theory of stroke volume determination by thoracic electrical bioimpedance. AACN Clin Issues 1999; 10 (3): 385–399.
Summers RL, Shoemaker WC, Peacock WF, Ander DS, Coleman TG . Bench to bedside: electrophysiologic and clinical principles of noninvasive hemodynamic monitoring using impedance cardiography. Acad Emerg Med 2003; 10 (6): 669–680.
Osthaus WA, Huber D, Beck C, Winterhalter M, Boethig D, Wessel A et al. Comparison of electrical velocimetry and transpulmonary thermodilution for measuring cardiac output in piglets. Paediatr Anaesth 2007; 17 (8): 749–755.
Zoremba N, Bickenbach J, Krauss B, Rossaint R, Kuhlen R, Schalte G . Comparison of electrical velocimetry and thermodilution techniques for the measurement of cardiac output. Acta Anaesthesiol Scand 2007; 51 (10): 1314–1319.
Tomaske M, Knirsch W, Kretschmar O, Woitzek K, Balmer C, Schmitz A et al. Cardiac output measurement in children: comparison of Aesculon cardiac output monitor and thermodilution. Br J Anaesth 2008; 100 (4): 517–520.
Norozi K, Beck C, Osthaus WA, Wille I, Wessel A, Bertram H . Electrical velocimetry for measuring cardiac output in children with congenital heart disease. Br J Anaesth 2008; 100 (1): 88–94.
Schubert S, Schmitz T, Weiss M, Nagdyman N, Huebler M, Alexi-Meskishvill V et al. Continuous, non-invasive techniques to determine cardiac output in children after cardiac surgery: evaluation of transesophageal Doppler and electric velocimetry. J Clin Monit Comput 2008; 22 (4): 299–307.
Rauch R, Welisch E, Lansdell N, Burrill E, Jones J, Robinson T et al. Non-invasive measurement of cardiac output in obese children and adolescents: comparison of electrical cardiometry and transthoracic Doppler echocardiography. J Clin Monit Comput 2013; 27 (2): 187–193.
Wong J, Agus MD, Steil G . Cardiac parameters in children recovered from acute illness as measured by electrical cardiometry and comparisons to the literature. J Clin Monit Comput 2013; 27 (1): 81–91.
Noori S, Drabu B, Soleymani S, Seri I . Continuous non-invasive cardiac output measurements in the neonate by electrical velocimetry: a comparison with echocardiography. Arch Dis Child Fetal Neonatal Ed 2012; 97 (5): F340–F343.
Grollmuss O, Demontoux S, Capderou A, Serraf A, Belli E . Electrical velocimetry as a tool for measuring cardiac output in small infants after heart surgery. Intens Care Med 2012; 38 (6): 1032–1039.
Grollmuss O, Gonzalez P . Non-invasive cardiac output measurement in low and very low birth weight infants: a method comparison. Front Pediatr 2014; 2: 16.
Torigoe T, Sato S, Nagayama Y, Sato T, Yamazaki H . Influence of patent ductus arteriosus and ventilators on electrical velocimetry for measuring cardiac output in very-low/low birth weight infants. J Perinatol 2015; 35 (7): 485–489.
Song R, Rich W, Kim JH, Finer NN, Katheria AC . The use of electrical cardiometry for continuous cardiac output monitoring in preterm neonates: a validation study. Am J Perinatol 2014; 31 (12): 1105–1110.
Blohm ME, Obrecht D, Hartwich J, Mueller GC, Kersten JF, Weli J et al. Impedance cardiography (electrical velocimetry) and transthoracic echocardiography for non-invasive cardiac output monitoring in pediatric intensive care patients: a prospective single-center observational study. Crit Care 2014; 18 (6): 603.
Lien R, Hsu KH, Chu JJ, Chang YS . Hemodynamic alterations recorded by electrical cardiometry during ligation of ductus arteriosus in preterm infants. Eur J Pediatr 2015; 174 (4): 543–550.
Cote CJ, Sui J, Anderson TA, Bhattacharya ST, Shank ES, Tuason PM et al. Continuous noninvasive cardiac output in children: is this the next generation of operating room monitors? Initial experience in 402 pediatric patients. Paediatr Anaesth 2015; 25 (2): 150–159.
Zubrow AB, Hulman S, Kushner H, Falkner B . Determinants of blood pressure in infants admitted to neonatal intensive care units: a prospective multicenter study. Philadelphia Neonatal Blood Pressure Study Group. J Perinatol 1995; 15 (6): 470–479.
Schneider DJ, Moore JW . Patent ductus arteriosus. Circulation 2006; 114 (17): 1873–1882.
Pinsky MR . Why measure cardiac output? Crit Care 2003; 7 (2): 114–116.
Engoren M, Barbee D . Comparison of cardiac output determined by bioimpedance, thermodilution, and the Fick method. Am J Crit Care 2005; 14 (1): 40–45.
Moritz B, Fritz M, Mann C, Simma B . Nasal continuous positive airway pressure (n-CPAP) does not change cardiac output in preterm infants. Am J Perinatol 2008; 25 (2): 105–109.
Luecke T, Pelosi P . Clinical review: positive end-expiratory pressure and cardiac output. Crit Care 2005; 9 (6): 607–621.
Reuter DA, Felbinger TW, Schmidt C, Kilger E, Goedje O, Lamm P et al. Stroke volume variations for assessment of cardiac responsiveness to volume loading in mechanically ventilated patients after cardiac surgery. Intens Care Med 2002; 28 (4): 392–398.
Gillebert TC, Van de Veire N, De Buyzere ML, De Sutter J . Time intervals and global cardiac function. Use and limitations. Eur Heart J 2004; 25 (24): 2185–2186.
Acknowledgements
This research received no specific grant from any funding agency.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Competing interests
The authors declare no conflict of interest.
Rights and permissions
About this article
Cite this article
Hsu, KH., Wu, TW., Wang, YC. et al. Hemodynamic reference for neonates of different age and weight: a pilot study with electrical cardiometry. J Perinatol 36, 481–485 (2016). https://doi.org/10.1038/jp.2016.2
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1038/jp.2016.2
This article is cited by
-
Electrical Cardiometry during transition and short-term outcome in very preterm infants: a prospective observational study
European Journal of Pediatrics (2024)
-
Usefulness of the patient-specific contrast enhancement optimizer simulation software during the whole-body computed tomography angiography
Heart and Vessels (2022)
-
Bioreactance-derived haemodynamic parameters in the transitional phase in preterm neonates: a longitudinal study
Journal of Clinical Monitoring and Computing (2022)
-
Prone sleeping affects cardiovascular control in preterm infants in NICU
Pediatric Research (2021)
-
Baseline cardiac output and its alterations during ibuprofen treatment for patent ductus arteriosus in preterm infants
BMC Pediatrics (2019)