Introduction

The FP 7 silicone Ahmed glaucoma valve (AGV) (New World Medical, Rancho Cucamonga, CA, USA) is an aqueous drainage device that is increasingly used to lower intraocular pressure (IOP) in refractory glaucoma. The principle of the method is that the aqueous humour is drained through a tube to a reservoir around an end plate, from which it is reabsorbed by the surrounding tissue capsule. One unique and attractive feature of the AGV is its valve-like mechanism to prevent over-filtration and hypotony in the early postoperative phase when the capsule has not yet built up around the end plate.1 The valve is located between tube and end plate. If the pressure in the tube exceeds a certain value, two elastic membranes of the valve open and the pressure drops. Once the pressure in the tube falls below a certain value, the membranes reappose, thereby closing the valve. This mechanism is designed to ensure a minimal postoperative IOP of 8 mm Hg (personal communication with Mr Ahmed ESCRS, September 2008).

Despite the outflow restriction, postoperative hypotony and related complications, such as choroidal effusion or shallow anterior chamber, are regularly reported.2, 3 In vitro studies have shown a high variability of the opening and closing pressures needed to activate or deactivate the AGV, which may explain cases of hypotony after AGV implantation.4

The present study aimed to assess whether intraoperative measurement of the opening and closing pressure can be used to identify individual valves with an increased risk of over- or underfiltration.

Materials and methods

We performed a single-centre study of 30 consecutive cases of glaucoma surgery with implantation of an AGV between February 2011 and April 2012. The AGV implantation was not combined with any other procedure, and no viscoelastic substances were used. All surgeries were performed by the same surgeon (FB). The study was approved by the local ethics committee and each patient gave informed consent. We certify that all applicable institutional and governmental regulations concerning the ethical use of human volunteers were followed during this research.

Surgical procedure

The AGV was primed as required by the manufacturer, by gently flushing balanced salt solution through the tube and valve prior to implantation. Immediately after the priming procedure and just prior to implantation, we intraoperatively measured the opening and closing pressures of the specific AGV device to be implanted. This measurement was performed by connecting the cannula with the attached tube of the AGV to an infusion bottle of our phaco-vitrectomy machine (Oertli OS3, Oertli Instrumente AG, Berneck, Switzerland). To equilibrate the system, the bottle height had been adjusted previously to the height of the instrument table so that no flow was observed in the tubing in either direction. Then the tube was connected to the infusion bottle, and the air pressure in the infusion bottle was increased by 1-mm Hg steps using the air infusion pump of the phako-vitrectomy machine. At each pressure step, the system was allowed to equilibrate during a 5-s pause. The pressure was increased until detection of flow through the valve of the AGV, indicating AGV activation (opening pressure A). At this point, the pressure was reduced in the same stepwise manner until the flow stopped (closing pressure B). Next, the pressure was increased again until flow reappeared (re-opening pressure C). Finally, the AGV was implanted using a standardised procedure. The surgery was performed under general or parabulbar anaesthesia. All AGVs were implanted in the supratemporal quadrant. The end plate was sutured to the sclera at a measured distance of 8–10 mm posterior to the limbus. Anterior chamber entry was made using a 23-gauge needle. A long oblique tunnel was created to prevent loss of fluid through the needle track. After the tube was inserted, the entry site was checked for leakage using a triangular cellulose sponge. In no case a leakage was observed. The limbal portion of the tube was covered with a pericardium patch.

Postoperative IOP measurement and data analysis

Intraocular pressure was measured postoperatively on the same day, 6 h after the surgery, using a Goldman applanation tonometer. The differences between pressures A and B, C and B, and A and C were calculated. Means and standard deviations were computed. Low IOP was defined as any value below 5 mm Hg. Pressure values of AGV with postoperatively low IOP were compared with those without low IOP using a two-sample t-test with equal variances.

Univariate analyses were used to assess each of the AGV pressure parameters and the differences against low IOP (as a dependent variable). Cut-off values for the two opening pressures and the closing pressure were calculated using the receiver operating characteristics (ROC) curve. The areas under the ROC curves (aROC) were calculated using the Stata 11.1 statistics software package (Stata Corp LP, College Station, TX, USA).

The probabilities for low IOP were estimated using a univariate logistic regression model, where a low postoperative eye pressure was entered as the dependent variate and valvular opening pressure was the independent variate. These values were plotted against the valvular opening pressure. To assess the overall reliability of the pressure measurement, we also calculated the correlation coefficients for the pairs of opening pressure, closing pressure, and re-opening pressure with the postoperative IOP.

Results

The clinical characteristics and the results of the opening and closing pressure measurements of the 30 patients are displayed in Table 1. The first 10 patients are the cases with low postoperative IOP<5 mm Hg.

Table 1 Clinical characteristics and results of the opening and closing pressure measurements

Figure 1 presents all individual valve pressure parameters and the postoperative IOP. The opening pressure A, closing pressure B, and re-opening pressure C ranged from 5–30, 3–18, and 4–20 mm Hg, respectively. The means (SD) for these three pressure parameters were 18.4 (5.1), 8.3 (4.7), and 11.7 (4.8)mm Hg, respectively. The difference between the first opening pressure A and closing pressure B was 9.8 (4.8). The difference between the closing pressure B and the second opening pressure C was 3.3 (2.5).

Figure 1
figure 1

Opening pressure A, closing pressure B, re-opening pressure C, and postoperative intraocular pressure of each patient.

Early after AGV implantation, 10 patients (33.3%) had low IOP. Three patients in this group suffered from IOP-related complications, such as shallow anterior chamber and choroidal effusion. Two cases (patient no 5 and 9) were managed successfully with injection of viscoelastics into the anterior chamber; in the 3rd case (patient no 1), the AGV was changed on the first postoperative day because of very shallow anterior chamber.

The 10 AGVs that resulted in low IOP showed significantly different mean opening pressure A, closing pressure B, and re-opening pressure C compared with those measured in AGVs without low postoperative IOP (Table 2).

Table 2 Mean IOP values of the 10 AGVs that resulted in low postoperative IOP compared with the measured values for the 20 AGVs without low postoperative IOP

Low IOP was associated with opening pressure A (OR, 0.55; 95% confidence interval (CI), 0.35–0.87; P=0.010), re-opening pressure C (OR, 0.56; 95% CI, 0.37–0.84; P=0.006), and to a lesser extent closing pressure B (OR, 0.61; CI, 0.41–0.9; P=0.013); the corresponding aROCs were 0.91 (0.81–1.00), 0.93 (0.84–1.00), and 0.88 (0.76–1.00), respectively (Figure 2).

Figure 2
figure 2

Receiver operating characteristic (ROC) curves for the opening pressure (a), closing pressure (b), and re-opening pressure (c).

An opening pressure A value of ≤18 mm Hg had a sensitivity (10/10) of 100% (95% CI, 69.2–100) and a specificity (13/20) of 65.0% (95% CI, 40.8–84.6) for identifying AGV that resulted in low IOP. For the closing pressure B, a cut-off value of ≤10 mm Hg had a sensitivity (10/10) of 100% (95% CI, 69.2–100) and a specificity (9/20) of 45.0% (95% CI, 23.1–68.5). For the re-opening pressure C, a cut-off value of ≤11 mm Hg had a sensitivity (10/10) of 100% (69.2–100) and a specificity (15/20) of 75.0% (50.9–91.3).

The correlation coefficients between postoperative IOP and the three pressure parameters were 0.73 for opening pressure A, 0.78 for closing pressure B (Figure 3), and 0.75 for re-opening pressure C.

Figure 3
figure 3

Correlation between closing pressure B and postoperative IOP.

Discussion

In the preoperative assessment of silicone AGV, we observed a large variability of opening, closing, and re-opening pressures. Re-opening pressure C with a cut-off value of 11 mm Hg was the most helpful parameter in identifying all AGVs that resulted in low IOP, while closing pressure B showed the best correlation between measured pressure values and postoperative IOP.

We also observed substantial variations in the differences between the opening and closing pressures. Thus, even if the pressure level leading to activation of the AGV is known, it is not possible to predict the pressure value at which the valve will stop the drainage. Furthermore, the second activation of the valve immediately after it had closed required less pressure, indicating that the pressure range over which one single AGV is active is also variable.

These results have an impact on the clinical use of the AGV, as they raise questions about the reliability of the valve mechanism as well as the priming procedure. A simple test of the specific AGV would be helpful to provide the surgeon with reliable information about its postoperative valve function. The herein described test procedure allows the prediction of situations with low postoperative IOP. In eyes that are prone to develop early postoperative hypotony, appropriate preventive measures can be taken, including exchange of the AGV before implantation.

It is important to note that low IOP does not automatically result in complications such as choroidal detachment or a shallow anterior chamber. Risk factors for a shallow anterior chamber after implantation of a glaucoma drainage device include myopia, small eyes, younger age, and fewer previous intraocular surgery.5, 6 In such cases, we recommend discarding the AGV if the opening pressure is 18.0 mm Hg or less.

Our data correlate well with the relatively high rate of hypotony reported in literature after implantation of silicone AGV, ranging from 1.8–49%.3, 7, 8 Another study reported hypotony-associated complications at rates of 19% (shallow anterior chamber) and 15% (choroidal effusion).9 Interestingly, our results confirm the findings of a study where silicone AGVs were tested in vitro; Moss and Trope4 reported that three out of six tested AGVs showed a very low closing pressure, ranging from 1.4 to 3.5 mm Hg.

Our study has several limitations. The test setting included subjective judging of the valve characteristics by the surgeon, and may have a limited accuracy. It is possible that loss of fluid along the tube track or low aqueous production could explain low postoperative IOP. However, the operation technique using a long oblique tunnel through which the tube enters the anterior chamber reduces the risk of leakage along the tube. Furthermore, it has been shown that flow characteristics might change in dynamic or steady-state flow conditions,10 and an in vitro study showed nonlinear resistance to flow through an AGV.11 Therefore, the test settings may not reflect in vivo conditions. However, the good correlation of the measured pressure parameters with the postoperative IOP indicates that the measurement is useful and reliable.

The reasons for the observed variability of the opening and closing pressures and for the highly variable pressure range over which an AGV is active are unclear. Therefore, the valve function as well the priming process of the AGV should be the focus of further research. It is possible that a standardised priming procedure could lead to more predictable valve function.

The valve function may not only play an important role in the early postoperative phase, but may also impact the long-term success of the glaucoma surgery. Like the surface structure and the shape of the AGV,12 the valve function defining the amount of drained aqueous accessing the sub-Tenon tissue may also influence the wound-healing response and the formation of an inflammatory capsule around the AGV.

In conclusion, pre-implantation assessment of the valve function is recommended. Our results showed that each of the three pressure parameters could identify AGV’s that resulted in low postoperative IOP with a cut-off value of 18 mm Hg for the opening pressure, 10 mm Hg for the closing pressure, and 11 mm Hg for the re-opening pressure. These threshold values should trigger an intraoperative exchange of AGV in cases prone to hypotony.