Introduction

Spectroscopy is a well-established tool used in the analysis of a material through identification of the spectral signatures of its constituents. Spectral imaging extends the usefulness of spectroscopy by combining spectroscopy with imaging, thereby providing both spectral and spatial information. Spectral imaging requires sensitive detectors and powerful computers to enable fast processing of images. Its use has expanded from remote sensing for both civilian and military purposes, to laboratory-based applications for biological imaging, such as cellular spectral imaging and oximetry. Spectral imaging systems record a stack of monochromatic images onto a two-dimensional detector array, such as a charge-coupled device (CCD); multiple images are collected over multiple wavelengths forming a ‘spectral data cube’. Spectral imaging systems incorporate a range of strategies to enable collection of the multiple monochromatic images. There has been widespread application of spectral imaging systems in applications ranging from cytogenetics and1, 2, 3, 4 pathology5 to oncology.6, 7

The use of spectral imaging to perform blood oximetry, exploiting the different spectral characteristics of oxygenated (HbO2) vs deoxygenated haemoglobin (Hb) has been described in a variety of clinical applications, such as assessing tissue perfusion,8, 9, 10 microvascular disease in diabetes,11 and sickle cell anaemia.12

Principles of blood oximetry

The development of the photoelectric cell enabled the quantification of light absorption, which contributed to the development of the spectrophotometer. As a result, the spectrophotometric study of the extinction coefficients (a measurement of how much light is absorbed by a substance at a given wavelength) of reduced Hb and oxyhaemoglobin was made possible; these measurements are central to the technique of oximetry. The first studies of the spectra of undiluted haemolysed and whole blood were undertaken by Drabkin13, 14, who demonstrated the applicability of the Lambert–Beer law in determining the OS of blood. The Lambert–Beer law, in transmission blood oximetry, assumes that for any given wavelength of light, its absorption is dependent on the extinction coefficient of the blood solution (ɛ), its concentration or haematocrit (c), and the distance (d) the light has to travel through the solution (path length):

where IT is the intensity of light transmitted through a solution and Io is the intensity of incident light.

The optical density (OD) of a solution is a measure of its attenuation, and is defined by:

Therefore, the OD of a solution of blood is a function of its extinction coefficient, concentration, and the path length the light has to travel. The main absorbing component of human blood is Hb, therefore, its extinction coefficient is assumed to be equal to that of Hb, although this is an approximation as it neglects scattering by the structure of red blood cells.

Kramer and Elam15 were able to demonstrate the linear relationship between the OS of blood and the absorption of red and infrared wavelengths of light. This finding was pivotal in the development of clinical oximeters. Matthes and Gross16, 17 were the first to demonstrate the use of an isobestic wavelength and an oxygen-sensitive wavelength in the red and infrared regions to compensate for variables, such as the haematocrit, path length, and light intensity. This technique of measuring the blood OS using an isobestic and an oxygen-sensitive wavelength is the basis of clinical dual-wavelength oximetry used today. The integration of optical plethsmography (analysis of the pulsatile component of the arterial cycle) with oximetry enabled the development of pulse oximeters, which are capable of measuring the arterial OS by isolating the arterial signal from the venous, bone, and tissue components.18 The calibration of pulse oximeters is performed using data acquired from normal subjects during induced hypoxia. Unsurprisingly, these calibration tests were carried out at arterial blood gas saturations of greater than 70%, the lowest tolerable state of systemic oxygenation in humans. At OSs below 70%, the calibration has been estimated by linear interpolation, therefore, the accuracy below this level is questionable.

Principles of retinal oximetry

Retinal oximetry requires an imaging system, which captures images of the retina using multiple wavelengths of light. Image analysis of this spectral image allows the OD of the retinal vessels to be estimated at several discrete wavelengths. The OD at a retinal location is obtained by calculating the ratio of the measured reflected light intensity adjacent to the target retinal vessel relative to the reflected light intensity at the centre of the retinal vessel:

where IV and IR are the intensity of light reflected from the retinal vessel and adjacent retina, respectively. The OD of the retinal vessels at various wavelengths can subsequently be used to calculate the OS (Figure 1).

Figure 1
figure 1

The molar extinction coefficients of deoxyhaemoglobin (Hb) and oxyhaemoglobin (HbO2) derivatives as a function of wavelength.

Various approaches have been undertaken to non-invasively measure the OS of blood in retinal blood vessels. Gloster19 measured changes in choroidal reflectometry with altered oxygenation. Laing20 developed a ‘choroidal eye oximeter’ to measure the OS of choroidal blood. Hickam et al21 described a photographic method to estimate the OS of the retinal veins in humans. Delori et al22 described a method for determining the OS in retinal vessels using a three-wavelength photoelectric oximeter; this comprised a fundus camera integrated with three interference filters, a photomultiplier, and a photocathode. The accuracy of the OS calculations was reported to be most accurate in the range of 50–100% OS, but in the range between 0 and 50% there was a tendency towards overestimation of the OS. Delori et al23 later applied the three-wavelength oximetry technique in conjunction with laser Doppler velocimetry to study the effects of neurogenic optic atrophy with associated inner retinal degeneration on oxygen delivery to the retina. In eyes with optic atrophy, the mean venous OS was reported to be significantly higher in both temporal (P=0.028) and nasal (P=0.001) veins compared with the fellow eyes. Beach et al24 developed a retinal imaging system using two wavelengths of light to measure the OS in the retinal vasculature of diabetic subjects without clinical evidence of diabetic retinopathy during normoglycaemia and hyperglycaemia. Schweitzer et al developed an imaging ophthalmospectrometer, which consisted of a modified fundus camera with an attached spectrograph. The instrument illuminated the retina with a small (40 μm × 1.5 mm) slit of light. In vivo measurements of retinal vessel OS were performed using the oximetry algorithm in 30 eyes of 28 healthy Caucasian subjects. The mean OS in 126 retinal arteries and 139 retinal veins were 92.2±4.1 and 57.9±9.9%, respectively. The central retinal venous OS was higher in patients with age-related macular degeneration compared with the controls.25 A study on the OS in the retinal vessels of patients with primary open-angle glaucoma (POAG) using the imaging ophthalmospectrometer were performed by Michelson et al. Retinal arteriolar OS in eyes with normal-tension POAG (89.7±5.4%) was significantly lower than the control group (P=0.006). Retinal venular OSs and arteriovenous differences in eyes with normal-tension POAG and high-tension POAG were not significantly different to the control eyes.26 The two-wavelength oximetry technique described by Beach et al24 was adopted by Stefánsson and co-workers who developed a retinal oximeter. Retinal vessel OSs of 16 normal eyes were compared at room air and during hyperoxia (100% OS). Significant differences in the retinal vessel OSs were demonstrated. In the transition from normoxia to hyperoxia, the mean retinal arteriolar OS increased from 96±9 to 101±8% (P=0.0027), and the retinal venules increased from 55±14 to 78±15% (P<0.0001). This experiment demonstrated that the retinal oximeter was capable of detecting oximetric changes in the retinal vessels during hyperoxia.27 A similar dual wavelength retinal oximeter to that reported by Hammer et al used the same concept of retinal oximetry described by Beach et al. The retinal oximeter comprised of a fundus camera (Carl Zeiss Meditec, Jena, Germany) with an attached dual-wavelength filter, which enabled the transmission of light at wavelengths of 548±10 and 610±10 nm (FWHM) to be recorded onto a colour CCD. The ODRs at the two wavelengths were used to calculate the OS. The mean (±SD) arteriolar and venular OSs were 98±10.1 and 65±11.7%, respectively. Inhalation of 100% oxygen increased the mean arteriolar and venular OS by 2 and 7%, respectively. The oximetry system was applied to study the OS in diabetic patients. Venular OS increased with the severity of diabetic retinopathy with a mean OS (±SD) of 69±7% in subjects with mild non-proliferative DR and 75±8% in subjects with proliferative DR.28

The application of a Fourier transform spectral imaging system to quantify the oximetric status of the retina with ischaemia was reported in a study of human subjects with retinal vein occlusions.29 More recently, the Fourier transform spectral retinal imaging system was used to compare the OS of the retinal tissues and blood vessels around the optic disc in normal eyes, and eyes with open-angle glaucoma.30 Overall, 15 eyes with low-tension glaucoma and 41 eyes with high-tension glaucoma were compared with 20 normal eyes. The average OS of the juxtapapillary retinal tissues in eyes with normal-tension glaucoma (78±8%) and high-tension glaucoma (82±10%) were reported to be significantly lower than in normal eyes (88±8%; P=0.002 and P=0.039, respectively).

Materials and methods

Hyperspectral fundus camera

The hyperspectral imaging system was built around a commercial mydriatic fundus camera (Canon CF-60Z, Tokyo, Japan), with a number of modifications to allow integration of a liquid crystal tunable filter (LCTF) into the optical path of the camera (Figure 2). This enabled illumination of the retina using a number of user-selected wavelengths of visible light. The LCTF chosen to be incorporated into the fundus camera was a CRI VariSpec (Woburn, MA, USA). The LCTF filter is able to transmit wavelengths between the visible and near-infrared region (420–720 nm) with a wavelength-dependent bandwidth of nominally 10 nm. Importantly, the LCTF is electronically tunable, enabling fast random access of the desired wavelength via a personal computer, contains no moving parts, and has an operating ambient temperature range between 10 and 40°C. The LCTF was integrated in an optimised position within the optical path of the incident light of the fundus camera. A low-noise CCD camera (Hamamatsu ORCA-ER, Hamamatsu Photonics KK, Hamamatsu City, Japan) was incorporated into the fundus camera to enable the electronic capture of retinal images. The CCD has a high resolution of approximately 1.3 million pixels and a dynamic range of 12 bits. The CCD is cooled to reduce dark current, noise, and thermal drift, and has a high-quantum efficiency in the visible and near-infrared wavelengths. A linear polariser is attached to the front of the CCD to reduce a fixed-pattern artifact caused by the specular reflection from the two surfaces of the objective lens within the fundus camera. Specular reflections from within the eye are also attenuated. A macro lens is inserted between the CCD and the fundus camera to maintain a field of view of approximately 40° onto the CCD detector chip with dimensions of 6.6 × 8.9 mm2 (Figure 2).

Figure 2
figure 2

Incorporation of the CCD and LCTF into the fundus camera.

The hyperspectral imaging system is controlled with a customised software programmed using LabView (National Instruments, Austin, TX, USA). This software was used to control the LCTF, trigger the xenon flash, and control the CCD camera to capture an image. The LCTF was controlled typically by changing the desired wavelengths in 10 nm increments or decrements. The xenon flash, engaged using the software, was integrated with the simultaneous capture of CCD images. A saturation histograph, live feed of the CCD recording, and image of a captured image were displayed on the software interface. The captured images were stored onto a personal computer in a 16-bit portable network graphics format.

Human imaging and spectral image analysis

We describe the use of hyperspectral fundus camera to acquire hyperspectral retinal images from healthy subjects and from an illustrative patient with a retinal branch artery occlusion, detailing the application of the image processing and analysis algorithms used to calculate the OS of the retinal vasculature. A total of 23 healthy subjects were recruited from the eye departments in Cheltenham and Gloucester General Hospitals. Written informed consent was obtained from all volunteers. All normal subjects were assessed by the study investigator. The best-corrected visual acuity (BCVA) was measured using a Snellen chart to record the distance vision. Intraocular pressures (IOPs) of both eyes were measured using Goldmann applanation tonometry. Slit-lamp biomicroscopy was used to assess the anterior segment. The pupils of the normal subjects were dilated with 1% tropicamide (Minims; Chauvin Pharmaceuticals, Romford, UK). Dilated fundoscopy was performed by slit-lamp biomicroscopy using a 78D lens. Oximetry measurements of the index finger were performed in all volunteers using a pulse oximeter (Biox 3740, Ohmeda, Louisville, CO, USA). Brachial blood pressure of all normal subjects was measured using an aneroid sphygmomanometer (Accoson 0342, AC Cossor & Son (Surgical) Ltd., Essex, UK).

Inclusion criteria were BCVA 6/9 or better, open anterior-chamber angle, absence of any ocular pathology, IOP <21 mm Hg, and normal optic-disc appearance. Exclusion criteria included history of intraocular surgery, ocular pathology, diabetes mellitus, stroke, hypertension, and glaucoma family history. The clinical details of the included normal subjects are summarised in Table 1.

Table 1 Summarised clinical data of the normal subjects included in the study

Acquisition of hyperspectral retinal images

Spectral images of the retina were acquired using wavelengths between 500 and 650 nm at 2 nm intervals. Eye fixation was maintained by instructing the subject to fixate on a red-fixation light with the contralateral eye. The average time to acquire a full data set of spectral images was approximately 10–15 min. Retinal images were obtained from 23 normal volunteers and from one patient with a branch retinal artery occlusion.

Results I

Qualitative features of the retinal blood vessels in spectral images of the retina

Spectral images of the retina revealed characteristic features of the retinal blood vessels. First, the retinal arterioles (Figure 3, red arrow heads) become less optically dense compared with the retinal venules (Figure 3, blue arrow heads) at wavelengths greater than 590 nm. This feature is consistent with the respective extinction coefficients of HbO2 and deoxyhaemoglobin; the OD of HbO2 is lower than that of deoxyhaemoglobin between 600 and 650 nm. Second, the macular branches of the retinal venules also appear to be less optically dense than the first-degree retinal venules, suggesting that the macular branches of the retinal venules appear to have relatively higher OSs than the first-degree venular vessels.

Figure 3
figure 3

The appearances of the arterioles and venules at nine selected wavelengths in the dark calibrated images of the retina in a normal subject. At wavelengths >600 nm, arterioles (red arrows) are less optically dense compared with the venules (blue arrows).

Results II

Processing of hyperspectral retinal images

The specific image processing algorithms that were applied to the retinal images are described in the following sections:

Image registration

The image registration method used a cross-correlation technique, which measures the degree of similarity between two images. A reference image is chosen, usually at 580 nm, in which the OD of the arterioles and venules are similar, and both demonstrate relatively high contrast relative to the retinal background. All of the other spectral retinal images are aligned to the coordinates of the reference image. The maximum cross-correlation within each image is searched for by rotating and translating the images, and comparing with the reference image. The cross-correlation techniques require features within each image to be similar in appearance, irrespective of the wavelength. However, the appearance of the spectral retinal images change with wavelength (Figure 3). Hence, a number of preprocessing steps were implemented during the image registration process. First, sequential spectral retinal images were acquired such that the main features of the retinal images (optic disc and large retinal vessels) were located in approximately at the same position in the image field. Second, the vascular network and the optic disc were used as features within each image, and enhanced to enable accurate cross-correlation between the reference retinal image and the sequential spectral images. These features were enhanced using customised spatial-frequency passband and edge detection (Laplacian or Gaussian) filters. Some image series could not be co-registered because of poor image quality, poor focus, and/or excessive movement between individual spectral images in the series; these image series were discarded.

Vessel detection and tracking

The coordinates of the centreline of the retinal vasculature were calculated by manually detecting the retinal blood vessels and applying a semi-automated vessel-tracking routine. The vessel-tracking technique is based on a Fast-marching algorithm, which searches for pixels between two points, with the lowest intensities, and finds the shortest path.31 To enable accurate tracking of the retinal vasculature, a reference retinal image (580 nm) was converted into an image, which highlighted the retinal blood vessels by assigning the lowest pixel values to the vessels in contrast with higher pixel values to the retinal background. Two points along the length of a given blood vessel segment within the reference image were then manually selected, and the vessel-tracking algorithm was applied. This process was repeated until the majority of the retinal vasculature was tracked. Accurate tracking for small arterioles and venules were often the most difficult to obtain, and therefore tracking of some of the smaller blood vessels were omitted because of vessel-tracking errors. The vessel-tracking algorithm enabled an estimate of the centreline coordinates of the retinal blood vessels.

Vessel profile extraction and calculation of the light transmission of the retinal vasculature

The centreline coordinates along each blood vessel were subsequently used to calculate the coordinates of a series of linear profiles orthogonal to the vessels. This was performed by calculating the perpendicular of the gradient between two successive points along the centre of a given blood vessel. The result is a series of linear profiles perpendicular to the blood vessels in the retinal vascular network (Figure 4). The length of the profile was manually selected depending on the size of the blood vessel, and was approximately 2–3 times the estimated width of the blood vessel. The intensity values along the length of each linear profile were extracted. This was repeated for all the spectral images such that for each linear profile in a given blood vessel, the intensity values along that profile were extracted for all wavelengths.

Figure 4
figure 4

Linear profiles perpendicular to the retinal blood vessels in an image of a normal right-eye retina.

Optical density estimation

Fitting algorithms were applied to each gray-scale intensity profile to estimate the intensity at the centre of the blood vessel (minima of the nonlinear curve) and intensity of the background adjacent to the blood vessel (gray-scale value of the linear fit at the centre of the blood vessel). The optical densities of a given point were estimated across the wavelength range (500–650 nm) to calculate the OD profile. This was repeated at all points along all the selected blood vessels.

Oxygen saturation calculation

This was performed by using an algorithm based on the Levenberg–Marquardt nonlinear fit to the complete set of OD (λ) for a given point on the tracked blood vessel in the equation,

where, CHb total is the total concentration of Hb, ɛoxy, and ɛde–oxy are the extinction coefficients of HbO2 and deoxygenated Hb, respectively, corrected for convolution with the spectral response of the LCTF; d is the vessel diameter, cOS is the calculated OS, η is the effective optical path-length contribution,38 and a is a scaling constant. The function accounts for the backscatter by blood cells. The algorithm estimates the free parameters (CHb Total,d, η, a, S) to provide an estimate of cOS. Oxygen saturation calculations were performed at all points along the tracked centreline of selected retinal blood vessels. Optimal OS results were attained when the spectral images between 556 and 640 nm were analysed. The OS calculations and corresponding coordinates of the centre of the tracked retinal blood vessels were used to generate oximetric pseudocolour maps of the retinal vasculature.

To enable further studies of the OS data, a semi-automated program was written in Mathematica (ver. 5.2; Wolfram Research, Inc., Champaign, IL, USA) to enable the extraction of calculated OS values at given points in the retinal arterioles and venules defined by the distance from the optic-disc margin. This was performed by determining the centre of the optic disc and its radius (r). Four additional circles with increasing radii (2r, 3r, 4r, 5r) about the point in the centre of the optic disc corresponded to the regions ½ disc diameter (dd), 1 dd, 1½ dd and 2 dd from the disc margin, respectively.

Oximetric maps of the retinal vasculature in normal subjects

Figure 5 shows pseudocolour images of the calculated OSs along the retinal arterioles and venules in three example normal subjects. It can be observed that the retinal arterioles generally have calculated OS between 70 and 100%. The calculated OSs of the retinal venules appear to be more varied. The mean OS calculation (±SD) of the temporal retinal arterioles and venules using a wavelength sequence between 556 and 650 nm were 104.3 (±16.7) and 34.8 (±17.8)%, respectively. The macular branches of the retinal venules generally appear to have higher calculated OSs compared with their respective main trunk first and second-degree retinal venules. Figure 6 shows box- and whisker plots of the OS of the (main trunk) temporal retinal arterioles and venules in normal subjects, with the median, range, and the minimum and maximum values displayed.

Figure 5
figure 5

Pseudocolour images of OS maps superimposed on the retinal arterioles (left) and retinal venules (right) in three normal subjects.

Figure 6
figure 6

A box-and-whisker plot of the OS of the temporal (superotemporal and inferotemporal) retinal arterioles and venules in normal subjects. The plots show the median, the interquartile range, 95% central range, and the minimum and maximum values.

These values changed very little with increasing eccentricity from the optic-disc margin. Figure 7 shows that the mean OS (and SEM) of the temporal retinal arterioles and venules at different (at the disc margin, ½ dd, 1 dd, 1½ dd and 2 dd) eccentricities from the optic-disc margin were not associated with significant changes in the OS; temporal retinal arterioles r2=0.071, P=0.66, and venules r2=0.002, P=0.95, respectively.

Figure 7
figure 7

Oxygen saturation calculations of the temporal (superotemporal and inferotemporal) arterioles (left) and venules (right) at points defined by the distance from the optic disc: at the disc margin (0.0), ½ disc diameter (0.5), 1 dd (1.0), 1½ dd (1.5) and 2 dd (2.0) from the disc margin. The mean OS (central black filled circle), the SEM (vertical lines), the best-fit line (solid black horizontal line), and the 95% confidence interval of the best-fit line (horizontal dashed lines) are shown in the plots.

Results III

Example of retinal vascular disease: branch retinal artery occlusion

Patient 1 was a 79-year-old man with a 25-day history of a left-inferior branch retinal artery occlusion (Figure 8a). The VA of the left eye was 6/18 at the time of imaging. The pseudocolour images of the OS calculations overlayed onto the retinal vessels (Figure 8b) indicate lower than normal OSs of the inferotemporal retinal arterioles, corresponding to the location of occluded retinal arteriole.

Figure 8
figure 8

(a) Colour fundus photograph of patient 1, showing pale inferotemporal retina, corresponding to region affected by inferotemporal retinal arteriole occlusion. (b) Pseudocolour oximetry map showing abnormally low OS within the affected inferotemporal retinal arteriole in contrast to the normal OS levels in the unaffected superotemporal arteriole. The corresponding inferotemporal retinal venule has a normal level of OS.

Discussion

The human retina has a dual blood supply to meet the high metabolic demands of the photoreceptors, ganglion cells, and the retinal pigment epithelium. The inner two-thirds of the retina is supplied from branches of the central retinal vessels, and the outer one-third of the retina is supplied by the choroid creating a watershed zone at the outer plexiform layer. The retinal circulation is characterised by a low blood flow rate (80 ml/min), but the choroid, in contrast, has a high level of flow (800 ml/min).32, 33 Choroidal arterio-venous OS difference is 3 vs (at least) 40% for the retinal circulation.34 This higher OS differences between arterial and venous blood in the retinal circulation offers the possibility of gaining insights into inner retinal health by the measurement of retinal vessel OS, and quantification of changes in arterio-venous OS, with physiological changes, for example, dark adaption and/or in response to retinal disease.

The accurate quantification of retinal oxygen concentration in retinal tissues, and the OS in retinal blood vessels, therefore, remains an important but challenging, goal. A number of different approaches have been reported in the literature. A significant proportion of retinal oxygenation studies have been focused on the use of oxygen-sensitive microelectrodes in studying the O2 tension (pO2) in the retina. Importantly, this technique is able to measure the pO2 gradients across the choroid, retina, and vitreous, which provide insights into the local metabolic state of the retina. Birol et al35 studied the distribution of foveal and perifoveal oxygen in six anaesthetised cynomolgus macaques using O2 microelectrodes. Two characteristic pO2 minima were identified in the fovea and perifovea at the level of the photoreceptor inner segments and at the vitreo–retinal interface. The invasiveness of this technique has restricted its use mainly in animal studies.36 Studies in human eyes have been restricted to measurements of vitreous pO2 during intraocular surgery.37, 38, 39

The potential value of the clinical measurement, but the invasive nature of the techniques described above, have ensured that non-invasive retinal vessel oximetry in humans remains an important aspiration. This report describes the design of a retinal camera, which has allowed successful acquisition of spectal images of the retina, and details the image processing and analysis techniques, which have been used to calculate the OS of the retinal vasculature in a small group of normal subjects and in an illustrative example of a patient with a branch retinal arterial occlusion. We have previously reported the validation of this spectral imaging technique using a model eye with artificial retinal vessels filled with human blood of known OS.40 A wavelength sequence of 556–650 nm was determined to be the most appropriate wavelength sequence to accurately calculate the OS of the retinal vasculature. This wavelength sequence incorporates the important features of the extinction coefficient curves of HbO2 (local minima and maxima) and deoxyhaemoglobin (local maxima) to enable the nonlinear fitting oximetry algorithm to accurately differentiate most OD profiles of blood with varying OSs.

A total of 14 normal volunteers were successfully imaged, and retinal oximetry values were obtained. Examples of the pseudocolour images of OS calculations, overlayed onto the retinal arterioles and venules in three example normal subjects are illustrated in Figure 5. Lower than expected OSs of the retinal arterioles can be observed frequently at locations where the arterioles are in close proximity to the larger venules. These errors are mostly due to inaccuracies of the vessel profile analysis; to analyse the light transmission of a vessel, linear profiles orthogonal to the tracked coordinates of the vessels are generated. The vessel profiles of tracked arterioles adjacent to the larger venules will crossover onto the spatial coordinates of the venules, resulting in the accidental ‘contamination’ of the OD calculation related to the arteriole compared with that associated with the venule (lower OS). Careful selection of appropriate points along the arterioles, at given distances from the optic-disc margin, was performed to reduce this tendency. Furthermore, OS calculations of small segments of the vessels were obtained for the segmentation analysis as opposed to large segments of the vessels to reduce the influence of spurious outlying results. Notwithstanding these precautions during analysis, physical trans-luminal diffusion of oxygen down its concentration gradient, from the arteriole to the venule, cannot be excluded as a potential additional factor. The cause of the variability of the OS in the retinal venules is unclear, and could be attributed to a number of factors, including variability in the illumination of the retina images (vignetting) and retinal background (variable pigmentation). Additionally, it could be explained by physiological variability in the function of the retina in the photopic conditions used to acquire the spectral retinal images, variable consumption of oxygen, and short-term changes in retinal blood flow. Analysis of the first- and second-degree temporal retinal arterioles at defined eccentricity from the optic-disc margin revealed a mean OS of 110.8 (±11.8)%. This is comparable to OS measurements of retinal arterioles in healthy subjects reported by previous authors. Correlation and linear regression analyses indicated that the OS of the first- and second-degree temporal retinal arterioles did not significantly change at varying eccentricity from the optic-disc margin of up to 2 dd. However, the mean OS of the first- and second-degree temporal retinal venules was 27.7 (±19.1)%, which is lower than the OS measurements of the retinal venules reported in previous studies. A possible explanation could be an error in the OS calculation caused by inaccurate curve fitting by the nonlinear algorithm to the OD profiles. However, this is unlikely as all the OD profiles included in the analysis were carefully selected and examined. Poor quality OD profiles were excluded from the analysis. The constant OS measurements of the retinal venules at different eccentricities from the optic-disc margin indicate consistency in the OS measurements. Previous retinal oximetry systems have reported retinal venous OSs ranging from 45 to 65%. Most of these systems rely on two wavelength oximetry, which have been reported to overestimate the true OS.34 A further possible explanation could be the wavelengths used by other workers,25 which ranged between 510 and 586 nm in 2-nm increments. At these wavelengths, the hyperspectral retinal imaging system described in this present study provided highly variable optical densities of the retinal vessels, which could possibly contribute to errors in the OS calculations.

The research work described has a number of limitations. Data is presented from a relatively small number of normal subjects. Data from nine normal subjects (40%) had to be excluded from the study because of poor image quality, resulting in poor OD calculations. Unfortunately this was unavoidable, as the current image acquisition process and subsequent analysis is cumbersome and demanding for both the volunteer and the operator. Acquisition and processing of the images to generate oximetry maps of the retinal vasculature of one eye took on average approximately 50 h of cumulative processing on a computer with a 2.5 GHz dual-core processor with 4 Gb of RAM. The processing and analysis steps are semi-automated, and in addition to this, further analysis had to be performed to calculate the OSs of the retinal vessels at defined points along the retinal vessels to give the mean OSs of the temporal retinal arterioles and venules. Furthermore, inaccuracies of the OD and OS calculations were frequently encountered in regions of the retinal images in which there was poor illumination of the retina, particularly towards the periphal nasal retina: calculated OSs beyond the physiological range (0–100%), negative values and some calculated OSs greater than 100%. As a result, analysis of segments of the retinal vasculature was limited to 2 dd, and analysis of the nasal retinal vasculature was not performed. In this study, truncation of the analysed wavelength sequence of available optical densities before oxygen calculations improved oximetry accuracy. The underlying cause of variability of the optical densities at wavelengths less than 556 nm is unclear. At these wavelengths, the retina appears dark in the spectral images. The reduced contrast between the retinal background and retinal vessels could be a plausible explanation for the variability in the OD calculations at these wavelengths. Although reducing the wavelength sequence has reduced the effects of this variability, the wavelengths between 556 and 650 nm are possibly still susceptible to variations in the retinal pigmentation. This was noticeable in the analysis of two normal subjects with highly pigmented retinas.

In summary, this study has described techniques to deliver retinal vessel OS calculations in normal subjects and in a retinal patient. Retinal arteriolar OS values were consistent with previous studies. Retinal venular OS values were lower than some reports on normal subjects in the literature, however, the values were comparable to those reported by Delori et al.22, 23 Retinal vessel oximetry using the described imaging and analysis techniques, especially using recently developed ‘snapshot’ spectral imaging refinements41 holds promise for accurate, practical, and non-invasive retinal oximetry measurements.