Abstract
Purpose
To evaluate the suitability of including both Heidelberg Retina Tomograph-I (HRT-I) and HRT-II examinations in the same longitudinal series for HRT topographic change analysis (TCA) and to evaluate parabolic error correction (PEC) to improve the agreement between HRT-I and HRT-II examinations.
Methods
A total of 66 eyes from the University of California San Diego Diagnostic Innovations in Glaucoma Study with baseline HRT-I and HRT-II examinations obtained on the same day and ≥3 HRT-II follow-up examinations were included. Two TCA analyses, HRT-I examination at baseline (HRT-I–mixed series) and HRT-II examination at baseline (HRT-II–only series) were compared. Agreement between the HRT-I–mixed and HRT-II–only series were estimated using Bland–Altman plots. Agreement was assessed: (1) using the current HRT software settings (PEC applied only to HRT-II–only series), and (2) modified HRT settings (PEC also applied to HRT-I–mixed series).
Results
With current HRT software settings, the HRT-I–mixed series significantly overestimated change locations (ie, red pixels) compared with the HRT-II–only series as indicated by statistically significant proportional biases in the Bland–Altman analysis. By applying PEC to HRT-I–mixed series there were no statistically significant biases in the TCA parameter estimates compared with the HRT-II–only series.
Conclusion
In some eyes, HRT-I and HRT-II baseline examinations are not interchangeable in TCA analysis without parabolic error correction. HRT-I–mixed series detected more changes characteristic of glaucoma when there were only minimal changes in the HRT-II–only series. Our results suggest that in the majority of cases, with PEC, HRT-I examinations may be included in a longitudinal series containing HRT-II examinations.
Similar content being viewed by others
Introduction
Detecting glaucomatous progression is one of the most important, yet challenging, aspects of disease management. This task generally is accomplished by examining serial visual fields, serial optic disk photographs and, more recently, serial structural imaging measurements. The latter technique is challenging, in part, because glaucoma is a slowly progressing disease requiring years of follow-up. During this time, structural imaging instruments typically undergo hardware and software improvements. Because patients are followed for many years, they often undergo testing using more than one instrument version. These versions ideally should be backwards compatible and provide comparable measurements.
The Heidelberg Retina Tomograph (HRT, Heidelberg Engineering, GmbH, Heidelberg, Germany) is a confocal scanning laser ophthalmoscope used for topographic assessment of the optic nerve head.1 HRT-I (classic) was introduced for clinical use in 1991 and HRT-II in 1999. The scans acquired using the HRT-I at 10° × 10° field of view and the HRT-II scans have the same transverse resolution of approximately 11 μm. Therefore, the topographies computed from the HRT-I and HRT-II scans are theoretically compatible and interchangeable. With current software, HRT-I scans are converted to HRT-II format and combined in a single longitudinal series with HRT-II scans to detect localized retinal height changes using topographic change analysis (TCA) and other change detection algorithms.2, 3, 4
We noticed that, in some eyes, TCA available in HRT software detected significantly larger number of locations with significant decrease in retinal height when using an HRT-I baseline examination compared to when using an HRT-II baseline examination conducted on the same day (for eg, see Figures 1a vs b and 2a vs b). In HRT software, a data normalization procedure corrects for any horizontal and vertical shifts, rotational and tilt misalignment, and differences in parabolic distortion (known as parabolic error correction (PEC), described in Subjects and methods section and in Appendix A) between baseline and each follow-up examinations in a longitudinal series. Because the field of view of HRT-I examinations (10° × 10°) is smaller than that of HRT-II examinations, PEC is not applied when an HRT-I examination is included in the longitudinal series. We hypothesized that some of the TCA differences observed in Figures 1b and 2b in contrast to Figures 1a and 2a, respectively, are because of the lack of PEC when using HRT-I examination as baseline.
In this study, we compare the TCA results when HRT-I and HRT-II examinations were used as baseline with HRT-II examinations as follow-up. Specifically, we first assess agreement in the TCA results with an HRT-I at baseline (HRT-I–mixed series) without PEC to same eyes with an HRT-II at baseline (HRT-II–only series) with PEC as in the current HRT software. We then change the data normalization procedure to also correct for differences in parabolic distortion when HRT-I examinations are used at baseline (as in Figures 1c and 2c).
Subjects and methods
Subjects
All participants in the University of California San Diego (UCSD) Diagnostic Innovations in Glaucoma Study (DIGS) with good quality HRT-I and HRT-II examinations acquired on the same baseline visit and with at least three good quality HRT-II follow-up examinations were included in the study. HRT examinations with mean pixel height standard deviation (MPHSD) <50 μm were considered to be of acceptable quality for analysis (after quality review according to standard protocols, HRT-I baseline examinations of five eyes in the HRT-I–mixed series and HRT-II baseline examinations of four eyes in the HRT-II–only series had MPHSD between 40 and 50 μm). The UCSD Institutional Review Board approved the study methodologies and all methods adhered to the Declaration of Helsinki guidelines for research in human subjects and the Health Insurance Portability and Accountability Act.
HRT instrumentation and data preparation
Because 10° HRT-I topographies have the same transverse resolution as 15° HRT-II topographies,1 HRT software can convert 10° HRT-I topographies to HRT-II format for analysis. During conversion, 256 × 256 HRT-I retinal height measurements are embedded in the center of a 384 × 384 pixel area and measurements outside the central 256 × 256 region are marked as invalid.
Two separate HRT databases were constructed to evaluate TCA results. In the HRT-II–only series database, only HRT-II examinations were included for analysis. In the HRT-I–mixed series database, the baseline HRT-II examination was replaced with an HRT-I examination acquired on the same day and all HRT-II follow-up examinations were retained as in the HRT-II database. HRT software versions used in this study were HeyEx 1.6.1.0, HRTS 3.1.2.5, and HRTImport 1.3.0.0 (Heidelberg Engineering, GmbH, Heidelberg, Germany).
Optic disk contour margins in both databases were manually drawn by the same person. Contour lines on the HRT-I–mixed series were adjusted to closely match that of the HRT-II–only series (relative differences in disk area were within ±5%). Topographies in relative-tilted coordinates were exported from the study databases.1
PEC for HRT follow-up topographies
For optic nerve head imaging, an optimal distance of 10 mm is recommended between HRT and the eye.1 When eyes are imaged at this optimal distance, the focal plane of HRT lies parallel to the retinal surface and HRT images are optimal with minimal distortion. When this optimal imaging distance is not maintained, the focal plane becomes distorted especially in the peripheral region (G Zinser and M Reutter, Heidelberg Engineering, Personal Communication, 2007; also refer to Appendix A).
Effects of the distance between HRT and the eye on the retinal focal plane are illustrated using ray-tracing diagrams in Figure 3. A wide-angle schematic eye model based on the Gullstrand-Le Grand model eye and a scanning beam of 2 mm diameter and 633 nm wavelength were used to construct the ray tracing diagrams using Optica 3 software (Barnhart Optical Research LLC, Urbana, IL, USA) in Mathematica 7 environment (Wolfram Research Inc, Champaign, IL, USA).5 We used 633 nm wavelength scanning beams instead of 670 nm scanning beams used in HRT because more accurate refractive indices of cornea (1.3747), aqueous humor (1.336), lens (1.4183), and vitreous humor (1.3347) were available at 633 nm for more accurate ray tracing.5
As illustrated in Figure 3, differences in the distance between HRT and the eye result in differences in degrees of distortion of the retinal focal plane that are more prominent in the peripheral region. In HRT, retinal images corresponding to these retinal focal planes are acquired through an imaging pinhole located at a plane conjugate to these retinal focal planes.6 This conjugate relationship between the retinal focal plane and the HRT pinhole located at the retinal conjugate plane is maintained at all working distances in HRT (G Zinser, personal communication, July 2010). Therefore, retinal measurements (at the pinhole) will have peripheral measurement distortions corresponding to the shape of these retinal focal planes. These differences in the retinal measurements result in differences in the curvature of the HRT retinal topographies that need to be corrected before analyzing HRT topographies for changes over time. Therefore, in addition to aligning topographies, HRT software also corrects for any difference in parabolic distortion of retinal measurements between baseline and follow-up examinations (parabolic error). Because parabolic distortion is more prominent in the peripheral region, HRT software corrects parabolic error only in 15° HRT-II topographies and not in 10° HRT-I topographies. Therefore, the current HRT software does not correct parabolic error when there is at least one HRT-I examination in a longitudinal series (as in the HRT-I–mixed series). The PEC procedure available in the HRT software (described in Appendix A) was implemented in MATLAB version 7.9 (The Mathworks Inc., Natick, MA, USA) and was separately applied to the HRT topographies in relative-tilted coordinates.
Topographic change analysis
HRT TCA is a three-way mixed effects ANOVA model for detecting localized (superpixel) retinal height changes and is commonly used for automated analysis of HRT topographies for detecting glaucomatous changes over time.2, 7, 8, 9, 10, 11, 12, 13 HRT TCA was implemented in MATLAB and superpixel change probabilities and mean difference topographies calculated. Four TCA change summary parameters (all units are in superpixels) were computed as described elsewhere:8 (1) total number of superpixel locations with significant decrease in retinal height (red pixels), (2) size of the largest cluster of red pixels, (3) total number of superpixel locations with significant increase in retinal height (green pixels), and (4) size of the large cluster of green pixels, all within the disk margin.
TCA agreement
Agreement between the HRT-I–mixed and HRT-II–only series was determined using the latest follow-up examination of each participant. For all TCA parameters, difference between the HRT-I–mixed and HRT-II–only series, means, and 95% confidence intervals (CIs) were estimated. Bland–Altman mean vs difference plots were generated and bias between the HRT-I–mixed and HRT-II–only series were estimated.14, 15 Observed bias is considered to be statistically significant only when the 95% CI of the bias does not include 0.
For quantitative evaluation of parabolic error in the HRT-I–mixed and HRT-II–only series, we estimated the mean parabolic error at a radius of 500 μm from the center of the parabolic error maps (Figure 7c) generated by the PEC procedure described in Appendix A.
Results
A total of 66 eyes from 66 UCSD DIGS participants were included. Participants ranged in age from 39.48 to 91.16 (median of 71.3) years. The median number of HRT follow-up examinations was five (range 4–7), representing a median of 5.27 years of follow-up. The mean spherical equivalent of refractive errors at baseline was −0.57 (95% CI: −1.08 to −0.07) diopters. There were no tilted optic disks in this study on the basis of the assessment of simultaneous stereophotographs by experienced certified graders. Demographics of the study participants and their baseline characteristics are presented in Table 1.
Agreement between the HRT-I–mixed and HRT-II–only series is summarized in Table 2. Using the current HRT software settings, the TCA red pixel parameters were significantly higher in the HRT-I–mixed series (without PEC) compared with the HRT-II–only series (with PEC). After applying PEC to the HRT-I–mixed series, there were no statistically significant differences between the HRT-I–mixed and HRT-II–only series.
Figure 4 shows the Bland–Altman plots for the TCA parameters, wherein parabolic error was corrected only in the HRT-II–only series as in the current HRT software. The estimated biases in Figure 4a and b had statistically significant positive slopes indicating that (1) the HRT-I–mixed series of the study participants significantly overestimated TCA red pixel parameters compared with the HRT-II–only series and (2) the difference between the HRT-I–mixed and HRT-II–only series increased in proportion to the TCA changes detected (ie, proportional bias). For the TCA green pixel parameters, observed biases and mean differences between the HRT-I–mixed and HRT-II–only series were of borderline statistical significance.
Figure 5 shows the Bland–Altman mean vs difference plots of the TCA parameters estimated after applying PEC to both the HRT-I–mixed and HRT-II–only series. For all TCA parameters, there was no statistically significant bias between the HRT-I–mixed and HRT-II–only series. PEC to the HRT-I–mixed series removed proportional biases in the red pixel parameter estimates (Figure 5a and b), and slightly increased the variability of differences in the green pixel parameters (Figure 5c and d).
Figure 6 shows the Bland–Altman mean vs difference plots of the TCA parameters estimated from the HRT-I–mixed and HRT-II–only series both without PEC. It can be observed that Figure 6 is very similar to Figure 4 indicating that HRT-I–mixed series without PEC has poor agreement with both HRT-II–only series with and without PEC. The mean parabolic error estimated at a radius of 500 μm from the center of the parabolic error maps (Figure 7c) was 27.42 (95% CI=23.44, 31.41) μm for the HRT-I–mixed series and −0.07 (95% CI=−0.56, 0.42) μm for the HRT-II–only series. Therefore, the parabolic error observed in the HRT-I–mixed series was significantly larger than in the HRT-II–only series further supporting the similarity observed between Figures 4 and 6.
For all of TCA parameters, there was no statistical significant association between the observed differences in the TCA parameter values and the differences in disk area (in pixels) between the HRT-I–mixed and HRT-II–only series (Kendall's Tau P-value >0.30).
Discussion
This analysis indicates that, with the current software settings, TCA detects more topographic locations with significant decrease in retinal height in longitudinal series with a baseline HRT-I examination (HRT-I–mixed series) without PEC compared with the same series with a baseline HRT-II examination (HRT-II–only series) with PEC. In some eyes, TCA detected changes in the HRT-I–mixed series suggest neuroretinal rim changes characteristic of glaucoma which are not apparent in the HRT-II–only series (Figures 1b and 2b). The differences and biases in the green pixel parameters observed between the HRT-I–mixed and HRT-II–only series were of borderline significance. Summary measures of both red and green pixel-based TCA parameters (Table 2) and their Bland–Altman plots (Figures 4, 5, 6) suggest that correcting parabolic error in the HRT-I–mixed series improves TCA agreement between the HRT-I–mixed and HRT-II–only series. Improved agreement observed for both red and green pixel-based TCA parameters indicates that the PEC procedure did not introduce significant number of erroneous green pixels while controlling red pixels and vice versa.
From the Bland–Altman plots in Figure 5, we observed the presence of a few outlying extreme differences even after applying the PEC to the HRT-I–mixed series. For example, in the largest cluster of red pixels plot in Figure 5b, differences >100 superpixels are present in five eyes even after applying PEC. There are several possible explanations for the differences in TCA results between the HRT-1–mixed series and HRT-II–only series. There may be differences between ‘eye data’ of the baseline HRT-I and HRT-II examinations, such as corneal curvature radius, astigmatic corrective lenses, glasses, and contact lenses that effect the magnification and scaling of the topographies and thus may significantly affect any topographic analysis. In this study, however, we observed no causal relationship between the eye data and the extreme outliers (data not shown). In addition, although HRT-I and HRT-II instruments have similar optical design, and acquire scans with compatible transverse resolution, there are several differences in their scan protocols. For example, in HRT-I, three separate optic disk scans are manually acquired in succession to constitute an optic disk examination, whereas three optic disk scans are automatically acquired in succession in HRT-II/3 instrument. Therefore, some differences are expected in the intraexam variability of HRT-I examinations and of HRT-II examinations. Moreover, the HRT-I acquires a fixed set of 32 optical section images irrespective of the scan depth, thus, has a variable axial resolution of 62–128 μm between optical sections. The HRT-II acquires a variable set of 16–64 optical sections and maintains a constant axial resolution of 62.5 μm between optical sections. It is expected that the differences in axial resolution between HRT-I and HRT-II instruments could be a source of variability between the topographies constructed from the HRT-I and HRT-II optic disk scans. There are other general improvements in the HRT-II instruments (such as tilted head rest, prescanning of the optic nerve head, faster scans, and so on.) that may also make it easier to take images at a consistent distance from the eye. The effects of these and other sources of measurement variability than can result in differences in the HRT-I and HRT-II images are difficult to characterize in this small cohort.
Currently, there are no studies on the repeatability of the HRT progression parameters for HRT-I–only longitudinal series and HRT-II–only longitudinal series (ie, how variable the progression parameters are if we replace the baseline examination of a longitudinal series comprised of only HRT-I examinations with another baseline HRT-I examination acquired on the same day; similarly, how variable the progression parameters are for the longitudinal series comprised of HRT-II examinations only). Therefore, it is not clear if the extreme outlying differences observed between the HRT-I–mixed series and HRT-II–only series are also likely between any two repeated HRT-I–only series or between any two repeated HRT-II–only series of an eye.
In addition, we modified the PEC calculations to include two additional strategies to improve the PEC procedure currently available in HRT software (data not shown). In brief, the modified PEC (1) excludes measurements within the optic disk to remove possible influence of optic disk measurements and any glaucomatous changes over time in eyes with disk diameter >1000 μm and (2) adaptively adjusts the dimensions of the peripheral retinal ring to correct for residual error during PEC. The modified procedure provided good agreement between HRT-I–mixed series and HRT-II–only series, similar to the current PEC procedure available in HRT software. By excluding optic disk measurements while estimating differences in parabolic distortion in a follow-up examination, the modified PEC procedure may reduce the possibility of correcting for changes in the follow-up examinations due to significant rim thinning or deepening of the optic cup.
Including an HRT-I examination at baseline with HRT-II follow-up examinations represents a common clinical scenario. In clinical practice, a TCA series may include several HRT-I examinations followed by HRT-II examinations because of software/hardware upgrades. The number of patients with HRT-I baselines and HRT-II follow-up is growing as patients continue to be examined with HRT-II and monitored using TCA analysis. It should be noted that the differences in TCA-detected retinal height decrease results between the HRT-I–mixed and HRT-II series did not occur in all eyes. In eyes with differences in TCA results, it is likely that the distance between the eye and HRT was different for the baseline HRT-I examinations and the follow-up HRT-II examinations, the condition that PEC was designed to correct. It is therefore difficult to estimate the proportion of eyes that can be expected to show the larger decrease in retinal height with an HRT-I as baseline compared with the HRT-II as baseline because the number can vary by clinic and operator. However, on the basis of significantly lower parabolic error observed in the HRT-II–only series, it can be inferred that HRT-II–only series may provide a more specific detection of glaucomatous changes.
We also investigated the possibility that the PEC procedure may be overcorrecting for differences in the HRT-II–only series thereby causing significant differences with the HRT-I–mixed series without PEC. Bland–Altman plots of TCA parameters estimated from the HRT-I–mixed and HRT-II–only series, both without PEC, also exhibited significant proportional biases in the TCA parameters (Figure 6). In addition, parabolic error present in the HRT-I–mixed series was significantly higher (27.42 μm) than the HRT-II–only series (−0.07 μm). Therefore, it is essential to correct parabolic error in the HRT-I–mixed series.
It is not readily apparent why HRT-II–only series have significantly lower parabolic error compared to the HRT-I–mixed series. In HRT-I topographies, peripheral retinal region outside 10° are not available. Therefore, during PEC, residual errors (calculated from retinal measurements outside the central 10°) are not corrected in the HRT-I–mixed series. We considered the absence of residual error correction in HRT-I–mixed series as a possible reason for the significantly higher parabolic error observed in the HRT-I–mixed series. To test this hypothesis, we calculated the parabolic error maps for several examples in the HRT-II–only series using only the measurements within the central 10° retinal topographic measurements and thus without correcting for the residual error. However, the parabolic errors were approximately the same for both with and without correcting for residual error in all of the HRT-II–only series examples tested. Further investigation in a controlled setting is necessary to understand the sources of higher parabolic error between HRT-I and HRT-II topographies.
Because the TCA disagreement observed between the HRT-I–mixed and HRT-II–only series are due at least in part to parabolic error, such disagreement is also expected when using other change detection algorithms.3, 4 In HRT, some of the rim, cup, and optic disk-related stereometric parameters are estimated using optic disk margin transferred from baseline to each follow-up after normalizing the topographies. Therefore, it is also important to evaluate agreement between the stereometric parameters of HRT-I and HRT-II examinations and the effects of PEC. This is studied in a separate work.16
One of the limitations of this study is the relatively small population size—a result of the strict eligibility criterion used for participant selection, where at least one good quality baseline HRT-I examination and one good quality HRT-II examination were required on the same day with at least three good quality HRT-II follow-up examinations for TCA analysis. A possible source of error in this agreement analysis is due to the necessity to manually draw optic disk contour lines separately on the HRT-I–mixed and HRT-II–only series.17 Possible contour line differences were minimized by closely replicating the contour line of the HRT-I–mixed series on to the HRT-II–only series by the same person (relative disk area differences were within ±5%). Moreover, we found no association between differences in disk size and differences in the TCA parameter estimates. Although this report focuses on the number of red and green pixels within the disk margin, we observed similar results for other TCA parameters including size of the largest red/green clusters relative to disk size, and area and volume of the red/green clusters within disk.
The PEC used in the HRT software is one of the data normalization procedures applied to HRT topography series before analyzing for changes over time. The other commonly known topography normalization procedures correct for horizontal and vertical shifts and rotational and tilt alignment among topographies in a longitudinal series. This study further highlights the importance of the data normalization procedures for preparing the high-resolution optical imaging datasets for progression analysis and their effects in the clinical diagnostic accuracy of the computational methods, such as TCA.
In summary, using the current HRT software, retinal height decreases in some eyes were overestimated by TCA when an HRT-I examination was used at baseline instead of an HRT-II examination. Application of PEC improved the agreement between the HRT-I–mixed and HRT-II–only series with few extreme outlying differences. Therefore, our results suggest that, with PEC, in the majority of cases, HRT-I examinations may be included in a longitudinal series containing HRT-II examinations.
References
Heidelberg Engineering, GmbH. Heidelberg Retina Tomograph: Operating Manual Software Version 3.0 2005.
Chauhan BC, Blanchard JW, Hamilton DC, LeBlanc RP . Technique for detecting serial topographic changes in the optic disc and peripapillary retina using scanning laser tomography. Invest Ophthalmol Vis Sci 2000; 41 (3): 775–782.
Patterson AJ, Garway-Heath DF, Strouthidis NG, Crabb DP . A new statistical approach for quantifying change in series of retinal and optic nerve head topography images. Invest Ophthalmol Vis Sci 2005; 46 (5): 1659–1667.
Balasubramanian M, Zabic S, Bowd C, Thompson HW, Wolenski P, Iyengar SS et al. A framework for detecting glaucomatous progression in the optic nerve head of an eye using proper orthogonal decomposition. IEEE Trans Inf Technol Biomed 2009; 13 (5): 781–793.
Escudero-Sanz I, Navarro R . Off-axis aberrations of a wide-angle schematic eye model. J Opt Soc Am A Opt Image Sci Vis 1999; 16 (8): 1881–1891.
Zinser G, Wijnaendts-van-Resandt RW, Ihrig C . Confocal Laser Scanning Microscopy for Ophthalmology. in SPIE Scanning Imaging. SPIE: Hamburg, Federal Republic of Germany, 1988.
Artes PH, Chauhan BC . Longitudinal changes in the visual field and optic disc in glaucoma. Prog Retin Eye Res 2005; 24 (3): 333–354.
Bowd C, Balasubramanian M, Weinreb RN, Vizzeri G, Alencar LM, O'Leary N et al. Performance of confocal scanning laser tomograph Topographic Change Analysis (TCA) for assessing glaucomatous progression. Invest Ophthalmol Vis Sci 2009; 50 (2): 691–701.
Chauhan BC, Hutchison DM, Artes PH, Caprioli J, Jonas JB, LeBlanc RP et al. Optic disc progression in glaucoma: comparison of confocal scanning laser tomography to optic disc photographs in a prospective study. Invest Ophthalmol Vis Sci 2009; 50 (4): 1682–1691.
Chauhan BC, McCormick TA, Nicolela MT, LeBlanc RP . Optic disc and visual field changes in a prospective longitudinal study of patients with glaucoma: comparison of scanning laser tomography with conventional perimetry and optic disc photography. Arch Ophthalmol 2001; 119 (10): 1492–1499.
Kalaboukhova L, Fridhammar V, Lindblom B . Glaucoma follow-up by the Heidelberg Retina Tomograph--new graphical analysis of optic disc topography changes. Graefes Arch Clin Exp Ophthalmol 2006; 244 (6): 654–662.
Kourkoutas D, Buys YM, Flanagan JG, Hatch WV, Balian C, Trope GE . Comparison of glaucoma progression evaluated with Heidelberg Retina Tomograph II versus optic nerve head stereophotographs. Can J Ophthalmol 2007; 42 (1): 82–88.
Vizzeri G, Weinreb RN, Martinez de la Casa JM, Alencar LM, Bowd C, Balasubramanian M et al. Clinicians agreement in establishing glaucomatous progression using the Heidelberg Retina Tomograph. Ophthalmology 2009; 116 (1): 14–24.
Bland JM, Altman DG . Measuring agreement in method comparison studies. Stat Methods Med Res 1999; 8 (2): 135–160.
Bland JM, Altman DG . Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986; 1 (8476): 307–310.
Balasubramanian M, Bowd C, Weinreb RN, Zangwill LM . Agreement between the Heidelberg Retina Tomograph (HRT) stereometric parameters estimated using HRT-I and HRT-II. Optom Vis Sci 2011; 88: (in press).
Garway-Heath DF, Poinoosawmy D, Wollstein G, Viswanathan A, Kamal D, Fontana L et al. Inter- and intraobserver variation in the analysis of optic disc images: comparison of the Heidelberg Retina Tomograph and computer assisted planimetry. Br J Ophthalmol 1999; 83 (6): 664–669.
Acknowledgements
We thank Gerhard Zinser, PhD, and Michael Reutter, PhD, Heidelberg Engineering, GmbH, Heidelberg, Germany for providing necessary support for conducting this study. This research was supported in part by Heidelberg Engineering, GmbH, Heidelberg, Germany, and the National Eye Institute Grant EY11008 (LMZ).
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Competing interests
Carl Zeiss Meditec: RNW (F, C), LMZ (F); Heidelberg Engineering: MB (F), LMZ (F), RNW (F); Lace Elettronica: CB (F), Paradigm: RNW (F); Topcon: RNW (F), LMZ (F); Optovue: LMZ (F).
Appendix A
Appendix A
Parabolic error correction applied only to Heidelberg Retina Tomograph-II (HRT-II)-only series (as in the current HRT software)
Algorithmic details presented here are based on an unpublished technical document from Heidelberg Engineering, GmbH, Heidelberg, Germany describing parabolic error correction procedure available in the HRT-3 software (HRTS module version 3.1.2.5) and through personal communication (Gerhard Zinser and Michael Reutter, December 2006 through January 2008).
The focal plane distortion because of the differences in the distance between HRT and the eye and the resulting retinal measurement distortion in the HRT topographies are approximately parabolic in shape as shown in Figure 3. Therefore, the curvatures of any two retinal topographies of an eye will be different if the same HRT–eye distance is not maintained between examinations. The resulting non-disease related curvature differences between any two retinal topographies of an eye are referred to as the parabolic error between the two examinations. Before conducting change analysis, such as HRT topographic change analysis, curvature differences (ie, parabolic error) among topographies should be removed to ensure that accurate difference images are constructed. Because the curvature differences are not purely parabolic, the parabolic error correction algorithm available in the HRT software updates an initial paraboloid fit for the curvature differences using an estimate of any residual error. The residual error reflects curvature differences present between the two topographies after the initial paraboloid fit, especially in the peripheral retinal region. Algorithmic details of the parabolic error correction algorithm are as follows.
Figure 8 shows the schematic layout of the parabolic error correction available in the current HRT software (HRTS module version 3.1.2.5) and Figure 7 (top row) shows an example parabolic error correction applied to an HRT-II baseline and follow-up examination pair. To characterize any differences in parabolic distortion between the baseline and each follow-up examination, a mean difference topography D(x, y) is calculated as the difference between the mean baseline topography and the mean follow-up topography. After excluding a central circular region of diameter 1000 μm (with center pixel coordinate: 193, 193) and the locations with invalid topographic height measurements (ie, topograph height = −32768 in the .RAW files), a paraboloid P(x, y)=a1+a2x+a3y+a4xy+a5x2+a6y2 is fit to describe the shape of the difference in the parabolic distortion between baseline and the follow-up examination. The mean difference topography after an initial parabolic error correction is given by D̂(x,y)=D(x,y)−P(x,y). Any residual difference in the parabolic distortion between baseline and the follow-up after the initial parabolic error correction is corrected using a residual error correction procedure as follows. The residual retinal height difference at a radius of Ravg (=175 pixels) from the center of the image between the baseline and follow-up examination is calculated from the parabolic error corrected retinal height differences D̂ at an angular interval of 1° within a peripheral ring of inner radius Rmin = 160 pixels and outer radius Rmax = 190 pixels, where Ravg = (Rmin + Rmax)/2. Thus, the residual retinal height difference (RD) is calculated as
The residual retinal height difference estimates are further spatially smoothed using a moving average filter with a spatial support of 23°. Using the angularly smoothed peripheral residual retinal height difference estimates (ie, RD), the final residual error correction matrix for the entire imaging area is calculated as follows.
Residual error correction matrix
Based on the residual error correction matrix estimate RC(r,ϕ), the initial parabolic error correction estimate P(x,y) is revised as P̂(x,y) by fitting a parabolid to the residual error corrected mean difference topography D(x,y)−RC(x,y).
Finally, differences in parabolic distortion between the baseline and the follow-up HRT examination are removed by subtracting the estimated total parabolic error from each of the 3 single topographies in the follow-up HRT examination.
Parabolic error correction applied to both HRT-I–mixed series and HRT-II–only series (not available in the current HRT software)
The parabolic error correction procedure described above for HRT-II–only series was also applied to all of the HRT-I–mixed series in the HRT-I–mixed database. The only difference is that because there is no valid pixels in the HRT-I topographies converted to HRT-II format in the peripheral retinal region (ie, at radius Ravg = 175 pixels), the residual error correction matrix RC = 0 and therefore, the revised paraboloid fit is same as the initial fit. The total parabolic error is given by the paraboloid fit P(x,y) for the mean height difference topography D(x,y). Figure 7 (bottom row) shows the application of parabolic error correction to an HRT-I–mixed series.
Rights and permissions
About this article
Cite this article
Balasubramanian, M., Bowd, C., Weinreb, R. et al. Agreement between Heidelberg Retina Tomograph-I and -II in detecting glaucomatous changes using topographic change analysis. Eye 25, 31–42 (2011). https://doi.org/10.1038/eye.2010.124
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1038/eye.2010.124