Introduction

Choroidal neovascularization (CNV) is a leading cause of loss of central vision in developed countries and may appear associated with different ocular disorders, the most common being age related macular degeneration (ARMD) and high myopia, and also after ocular inflammation and angioid streaks.1 New advances in the treatment of CNV, such as photodynamic therapy and transpupillary thermotherapy, have added new value to the determination of the activity of CNV through measurement of retinal thickness at the posterior pole.2, 3

Retinal thickness analyser (RTA) is a new device for the study of the topography of the posterior pole of the eye. It projects 15-μm-wide slits of green light of 534 nm, and records and digitizes light scattered back from the retina, and then localizing both interfaces and measuring the distance between them. These two peaks are later detected by a computer and the difference between them is measured, using this measure to map the posterior pole and calculate its thickness.4

Optical coherence tomography (OCT) provides high-resolution cross-sectional images of the retina permitting visualization of abnormalities in the retinal structure. It projects an infrared light of 820 nm, which produces high-resolution (axial 10 μm, transversal 20 μm) interferometry cross-sectional images of the retina.5

In the present study, we have used the RTA and the new version of OCT (OCT3) to study changes in the vitreoretinal and chorioretinal surface in eyes with subfoveal CNV.

Patients, materials and methods

We have used the commercially available latest version of OCT (OCT3, Zeiss-Humphrey, San Leandro, CA, USA) and of RTA (Talia Technology Inc., Israel).

In all, 11 patients with macular CNV associated to ARMD were studied using RTA and OCT3 by experienced operators. Measures were taken consecutively on the same day in order to make them comparable.

RTA was used to perform scans of a 6000 × 6000 μm2 area of the posterior pole centred in the fixation point (from 3000 μm temporal to 3000 nasal to fovea, and from 3000 μm superior to 3000 μm inferior to fovea), using the posterior pole and peripapillary thickness function.

OCT3 performed 6 mm scans centred in the fovea (a circle 6000 μm in diameter) using the fast macular scan function. Fixation was controlled by the video image of the central retina.

We have compared the OCT3 results of macular thickness in patients with CNV, with those of 40 healthy eyes (mean age 47.7 years range 20–78 years; SD 18.2) from our database, which showed only mild refractive defects (−4 to +2 D). For RTA we have used published data of healthy eyes.6 Linear correlation test was performed in order to compare the results by both procedures.

The results were compared in terms of absolute and relative thickness measured, and limitations and advantages of the procedures.

Results

Two patients could not be studied with RTA because of lack of central fixation. One patient could not have RTA performed because of high myopia (−12 D). Five men and three women were fully studied, one eye in each patient, five right, and three left eyes. The mean age of patients was 73.5±9.6 years (range 56–87 years).

OCT provided good-quality images in all cases; meanwhile, RTA images showed partial lack of information in two cases (25%).

The mean foveal thickness (MFT) as measured with OCT was 289.9±92.1 μm and 207.7±60.8 μm as measured with RTA (r=0.954 ). The mean perifoveal thickness (MPT) was 293.8±46.3 μm according to OCT and 200.8±61.3 μm according to RTA (r=0.920). The maximal perifoveal thickness (MT) measured by OCT was 335±104.0 μm and 316±86.8 μm by RTA (r=0.707). Differences between both systems were more important in those patients with partial lack of information (Table 1, Figures 1, 2 and 3).

Table 1 Macular thickness (μm) measured by OCT and RTA
Figure 1
figure 1

Scatter of MFT, r=0.954.

Figure 2
figure 2

Scatter of MPT, r=0.920.

Figure 3
figure 3

Scatter of MT, r=0.707.

Graphic topographic information was clear, intuitive, and self-explanatory in both instruments, although it was more marked with RTA (Figures 4a and b).

Figure 4
figure 4

(a) RTA Topographic image shows highs and deeps of retinal surface in shapes and colours (case 2). (b) OCT topographic image in the same eye shows similar colour profile of the vitreoretinal interface. (c) CNV activity is proved by the presence of fluid deep in the retina (arrow). (d) This OCT image shows the presence of CNV with subretinal fibrosis, although no fluid accumulation can be seen.

OCT also provided cross-sectional images of the retina, which helped in determining whether the increase in retinal thickness was due to the presence of CNV itself or to the presence of accompanying fluid. Six patients showed active CNV expressed by the presence of fluid under the retina, and two showed lack of fluid by retinal thickening caused by the presence of CNV alone.

Case 1 An 85-year-old lady with ARMD and good central fixation capability showed an MFT of 217.6 μm with RTA and 302 μm with OCT. Topographic images showed similar location of retinal thickening in both cases, mainly nasal and inferior to fixation. Cross-sectional images by OCT, although, showed the presence of fluid under the retina proving the activity of CNV.

Case 2 A 73-year-old man with ARMD showed an MFT of 226.3 μm with RTA and 375 μm with OCT. MPT was 224.2 μm with RTA and 363.0 μm with OCT. Central fixation was good. Topographic images showed in both cases an almost doughnut-shaped elevation of the posterior pole corresponding to CNV. OCT images showed that retinal engorgement was caused by CNV with fluid accumulation under the retina (Figure 4c).

Case 3 A 77-year-old man with ARMD and poor central fixation showed an MFT of 181.6 μm with RTA and 202 μm with OCT. MPT was 182.3 μm with RTA and 241.5 μm with OCT. Owing to poor central fixation RTA measures skipped over 75% of posterior pole giving way to misleading values. Topographic images showed retinal moderate thickening in the inferior nasal area with both techniques. OCT also indicated the presence of an atrophic CNV that produced no fluid around it, justifying the moderate thickening and poor fixation capability (Figure 4d).

Discussion

Retinal thickness in patients with CNV was increased in all measurements performed with both instruments. Thickness values were higher for OCT in absolute values but not as much when compared to values in healthy individuals. For RTA, MFT was 136.5% over healthy individuals (152.1 μm) and MPT was 114.7% (175.0 μm). For OCT, they were 132.4% over healthy individuals (219.8 μm) and 108% (272.5 μm), respectively. These results confirm those previously published on correlation between measures in both systems for healthy patients.7

Both instruments have previously proved to offer repetitive results concerning retinal pathology, especially for diabetic macular oedema.5, 6, 8

The differences found between both instruments may be related to several factors. First, the area considered by each device when measuring foveal or perifoveal thickness is not the same. RTA MFT takes into account the thickness of a circle of a 2000 μm radius centred on the fixation point, whereas MPT is defined as the thickness in doughnut-shaped area around the fovea with an internal radius of 600 μm and external radius of 2500 μm.6 OCT defines MFT as that of a circular area centred in the fixation point with a 1000 μm radius, and for MPT it considers a doughnut-shaped area around the fovea with an internal radius of 500 μm and external radius of 1500 μm.

Second, both systems take into account the distance between vitreoretinal and chorioretinal interfaces, by different procedures, using different wavelengths. Infrared light used by OCT allows imaging of structures placed deep in the retina in spite of the presence of vascular tissue such as CNV or haemorrhages. Measurements of retinal thickness are obtained by using computer image processing techniques in both techniques, although OCT also allows manual measuring at selected points of the retina. RTA may show CNV by the elevation on the vitreoretinal surface.7

And third, lack of information related to certain areas of the posterior pole due to insufficient capability of fixation in RTA can also lead to errors in the evaluation of mean values.

The time required for image acquisition in RTA is variable and longer than required for OCT, since it requires several steps with changes in fixation. RTA changes the location of the fixation point in order to measure different parts of the posterior pole, forcing the follow movement of the eye. Thus, it demands a good fixation capability for an accurate measure, since data are discarded when fixation has not been good enough. On the other hand, OCT makes a multiple scan in 1.92 s, 3.5 or 6 mm in diameter, of six different meridians centred in a chosen point. Although both techniques require good fixation, the need for changing the fixation in order to change the scanned field with RTA makes it more difficult for the patient with poor fixation capability. OCT can use either internal or external fixation, and the operator can move the scan area and centre it in the fovea even if the patient is not able to follow the light.

One would expect that the fixation would be less of a problem with the RTA since the patient only needs to fix for about 0.3 s to acquire a scan. However, it is necessary for the patient to fix five times since five separate scans are taken that are then brought into registration by the software (or manually, if necessary). By comparison, the OCT requires 1.92 s of continuous fixation to acquire the six scan lines. As five scans are acquired, the time from start to finish is longer for RTA even though the total time the retina is being scanned by the instrument is less for RTA than for OCT. The quality of the data depends on the duration required for each scan, not the total time the patient is seated at the instrument for measurement.

The RTA has quality control procedures to reject data that are not satisfactory. This is not the case for OCT, which could produce unsatisfactory data without notifying the operator. Yet the appearance of discordant measures will make the operator aware of this. Both systems allow visualization of the fundus and manual centring of the data acquisition by the operator, if necessary.

RTA samples a large area of posterior pole by slits separated by less than 200 μm. This information is used to calculate values in at least 1280 locations from which the topographic map is generated. In contrast, only six 6-mm lines of data are used to construct the topography with the OCT. However, no data are available for locations between the radial lines of data acquisition, and these are separated by as much as 1.5 mm. The gaps in data are most prominent far from the fovea. Accordingly, while the unmeasured locations allegedly have values determined by interpolation, in reality most of the values are essentially fabrication. Furthermore, the centre of the map is constructed on the assumption that all the lines pass through the centre of the fovea. This assumption may not be met in some patients due to the difficulty in sustaining fixation for the entire data acquisition time.

Both systems are much limited by media opacities such as lens or posterior capsule opacities and by pupillary diameter, offering poor quality images that cannot be measured either manually or by the software. In terms of this aspect we have found no difference between them.

RTA has been used in the assessment of retinal thickness in diabetic patients with and without diabetic retinopathy,6, 7, 8 and proved to be reproducible. Although RTA may be useful for measuring diffuse retinal thickening, such as that of diabetic retinopathy, the more localized thickening and the poor fixation often found in patients with CNV may make it less sensitive in these patients.

Both systems provide useful information in measuring retinal thickness in patients with posterior pole pathology. The limitations that RTA imposes due to fixation needs may work at the same time as a ‘quality filter’, improving the quality of the measures, and also reducing the number of patients who can benefit from this diagnostic procedure.

Cross-sectional images provided OCT, which gave important information to enable evaluation and to enable decision making in terms of patients with CNV, since they inform about the indepth composition of the thickening, especially the presence of fluid and the shape, thickness, type, and size of CNV.9

In conclusion, OCT and RTA may provide important information in the study and follow-up of patients with CNV, although thickness measures cannot be compared between both techniques or be considered in terms of absolute values since they are obtained by different means. The important relationship of RTA to fixation may make it less suitable for patients with foveal damage.